Magnetom Aera and 1 more
System, Nuclear Magnetic Resonance Imaging. 510(k) K101347, from Siemens Medical Solutions USA, Inc. (Malvern, PA). Product code LNH, class 2.
As of Nov 20, 2026, K101347 is an FDA 510(k) held by Siemens Medical Solutions USA, Inc., substantially equivalent on Oct 1, 2010 after 141 days, with 7 recalls tied to it in FDA's recall database.
- Review time: 141 days
- Recalls: 7
- Cited as a predicate by: 7 devices
The MAGNETOM systems described above are indicated for use as a magnetic resonance diagnostic device (MRDD) that produces transverse, sagittal, coronal and oblique cross sectional images, spectroscopic images and/or spectra, and that displays the internal structure and/or function of the head, body, or extremities.
- Magnetom Espree SystemK041112 · Siemens Medical Solutions USA, Inc. · Jul 21, 2004
- Magnetom VerioK072237 · Siemens Medical Solutions USA, Inc. · Oct 10, 2007
- Syngo Mr B17K082427 · Siemens Medical Solutions USA, Inc. · Nov 7, 2008
- MAGNETOM Vida, MAGNETOM Lumina, MAGNETOM Vida FitK192924 · Siemens Medical Solutions USA, Inc. · Mar 11, 2020
- Magnetom Sola, Magnetom Altea and Magnetom Sola FitK192496 · Siemens Medical Solutions USA, Inc. · Feb 28, 2020
- Imri 1.5T and Imri 3T SK133692 · Imris, Inc. · Feb 10, 2014
- Magnetom Avanto-Fit, Magnetom Skyra-FitK130885 · Siemens Medical Solutions USA, Inc. · May 17, 2013
- Magnetom SpectraK121160 · Siemens Medical Solutions USA, Inc. · Jul 16, 2012
- Syngo.Mr Spectro Engine, Syngo.Mr Spectro Svs, Syngo.Mr Spectro Csi, Syngo.Mr Spectro ExtensionK120315 · Siemens Medical Solutions USA, Inc. · Apr 13, 2012
- Magnetom Aera and Magnetom SkyraK111242 · Siemens Medical Solutions USA, Inc. · Nov 23, 2011
| Aug 28, 2025 | Z-2660-2025, Z-2655-2025There is a potential for an ice blockage to form or currently exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed venting paths, leading to a… | Open |
| Oct 17, 2016 | Z-1770-2017Due to a potential manufacturing error, a rare possibility exists for a sudden drop in table height of the Combi Dockable Table Neurosurgery. | Terminated |
| Apr 13, 2016 | Z-1598-2016The docking station top cover of the removable patient table may contain a magnetic metal plate. When docking/undocking the table, the magnetic top cover may become attracted to the magnet. | Terminated |
| Sep 12, 2012 | Z-2464-2012When positioning the Flex Large 4 Coil (part No. 8625761) off-center in left-right direction as it is used for example for hip and shoulder imaging and, in addition at an off-center position with respect to the… | Terminated |
| Jan 26, 2012 | Z-0940-2012During product monitoring firm has discovered that on a few systems the gradient cable connections did not meet their specifications after installation. Locally, the cable connections overheated, making servicing… | Terminated |
| Jun 20, 2011 | Z-2979-2011Firm discovered during product monitoring that on a few systems the gradient cable connections did not meet their specifications after installation. Locally, the cable connections overheated, making servicing… | Terminated |
- MAGNETOM Free.XLK262011 · Siemens Healthineers AG · Nov 5, 2026
- ECHELON SynergyK262902 · Fujifilm Corporation · Oct 9, 2026
- Vista AI ScanK260092 · Vista AI, Inc. · Oct 7, 2026
- SwiftSight-BodyCompositionK261612 · Airs Medical, Inc. · Oct 1, 2026
- APERTO Lucent MRI SystemK261283 · Fujifilm Corporation · Sep 16, 2026
- Embrace Vision Neonatal MRI SystemK261758 · Aspect Imaging, Ltd. · Sep 16, 2026
- MAGNETOM Free.Max; MAGNETOM Free.StarK260758 · Siemens Shenzhen Magnetic Resonance , Ltd. · Sep 1, 2026
- Swoop® Portable MR Imaging® SystemK262622 · Hyperfine, Inc. · Aug 19, 2026
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- May 13, 2010
- Review panel
- Radiology
- Regulation
- 21 CFR 892.1000
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files