Vasoview Hemopro 2 Endoscopic Vessel Harvesting System
Electrosurgical, Cutting & Coagulation & Accessories. 510(k) K101274, from Maquet Cardiovascular, LLC (San Jose, CA). Product code GEI, class 2.
As of Nov 20, 2026, K101274 is an FDA 510(k) held by Maquet Cardiovascular, LLC, substantially equivalent on Jun 11, 2010 after 36 days, with 10 recalls tied to it in FDA's recall database.
- Review time: 36 days
- Recalls: 10
- Cited as a predicate by: 3 devices
he 510(k) number has not been issued yet. Device name VASOVIEW HEMOPRO 2 Endoscopic Vessel Harvesting System Indications for Use The VASOVIEW HEMOPRO 2 System is indicated for use in minimally invasive surgery allowing access for vessel harvesting, and is primarily indicated for patients undergoing endoscopic surgery for arterial bypass.
- Guidant Vasoview Hemopro Endoscopic Vessel Harvesting System, Model VH-3000K052274 · Guidant Corporation · Sep 21, 2005
- Guidant Vasoview 6 Harvesting Cannula, Model VH-2000; Guidant Vasoview 6 Accessory Pack, Model VH-2004K041981 · Guidant Corporation · Aug 20, 2004
- Vasoview Hemopro 3 Endoscopic Vessel Harvesting System (VH-6000); Vasoview Hemopro 3 Endoscopic Vessel Harvesting…K251238 · Maquet Cardiovascular, LLC · May 30, 2025
- Vasoview Hemopro 3 Endoscopic Vessel Harvesting System (VH-6000), Vasoview Hemopro 3 Endoscopic Vessel Harvesting…K233879 · Maquet Cardiovascular, LLC · Feb 27, 2024
- VASOVIEW HEMOPRO Extension Cable, VASOVIEW HEMOPRO 2 Extension Cable, Bipolar Extension Cable, Fixed Distance, Active…K191947 · Maquet Cardiovascular, LLC · Sep 17, 2019
| Jun 2, 2026 | Z-2741-2026Getinge identified that Vasoview Hemopro 2 devices from Lot 3000547797 were exposed to an out-of-specification sterilization dose during a gamma sterilization run (Run ID 1183-28782A) performed by STERIS at its… | Open |
| Mar 20, 2025 | Z-1622-2025There have been at least 12 complaints from users as of March 3, 2025, reporting that they identified the ceramic C-Ring to be broken in half during use of the device. The ceramic C-Ring has a potential to break while… | Open |
| Mar 10, 2025 | Z-1577-2025Incorrect resistor utilized in the VH-3010 Power Supply, which may cause the jaws of the HemoPro 2 to not adequately heat up. | Open |
| Dec 9, 2024 | Z-0712-2025, Z-0711-2025Potential for two issues: 1. Bent or detached heater wire; 2. Silicone peeling or detaching from the Jaws of the Harvesting Tool | Open |
| May 31, 2024 | Z-2338-2024There were 27 complaints between March 22, 2024, and April 30, 2024, related to VH-3500 and VH-4000, reporting that the C-Ring wire was straightened (due to insufficient bend radius) and that the distance between C-Ring… | Open |
| May 17, 2024 | Z-2175-2024Potential for there to be a fluid ingress into the handle of the harvesting tool, which may lead to the device not providing cautery or providing cautery when not intended. | Open |
| Mar 6, 2023 | Z-1336-2023Some batches of product were not sterilized to their minimum sterilization specification. | Open |
| Dec 22, 2022 | Z-0995-2023Ink on the Vasoshield syringe Maquet logo may chip resulting in an unreasonable risk of harm to the patient. Potential harms may include foreign body reaction, coronary embolic event, and peripheral embolic event. | Open |
| Oct 30, 2013 | Z-0466-2014Maquet has received an increase in the number of complaints for "self-activated/remains activated/overheats" for the Hemopro VH-30000, "Hemopro 1" | Terminated |
- SP Electrocautery Device (SP18 Max);SP Electrocautery Device (SP20 Max)K262720 · Single Pass, Inc. · Oct 26, 2026
- SafeAir Electrosurgical Generator with Smoke Evacuation (0795-001-000);SafeAir Generator 3-Pedal Footswitch…K262386 · Stryker Instruments · Oct 20, 2026
- HARMONIC 1100 Shears (HAR1120); HARMONIC 1100 Shears (HAR1136)K254196 · Ethicon Endo-Surgery, LLC · Sep 17, 2026
- Neptune® SafeAir® 10ft Smoke Evacuation Pencil, Square Coated, 70mm Blade, Push Button Switch (0703-046-100);Neptune®…K261959 · Stryker Instruments · Sep 17, 2026
- LigaSure™ RAS Maryland Jaw Vessel Sealer/Divider, Nano-coatedK261070 · Covidien, LLC · Sep 15, 2026
- Potenza PrimeK254185 · Jeisys Medical Incorporated · Sep 11, 2026
- Ellian Enigma RF DeviceK262364 · Ellian Corp · Sep 2, 2026
- Scarlet Pro (Scarlet Pro)K253859 · VIOL Co., Ltd. · Aug 27, 2026
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional, third-party review
- Received
- May 6, 2010
- Review panel
- General, Plastic Surgery
- Regulation
- 21 CFR 878.4400
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files