Vasoview Hemopro 2 Endoscopic Vessel Harvesting System

Electrosurgical, Cutting & Coagulation & Accessories. 510(k) K101274, from Maquet Cardiovascular, LLC (San Jose, CA). Product code GEI, class 2.

As of Nov 20, 2026, K101274 is an FDA 510(k) held by Maquet Cardiovascular, LLC, substantially equivalent on Jun 11, 2010 after 36 days, with 10 recalls tied to it in FDA's recall database.

  • Review time: 36 days
  • Recalls: 10
  • Cited as a predicate by: 3 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

he 510(k) number has not been issued yet. Device name VASOVIEW HEMOPRO 2 Endoscopic Vessel Harvesting System Indications for Use The VASOVIEW HEMOPRO 2 System is indicated for use in minimally invasive surgery allowing access for vessel harvesting, and is primarily indicated for patients undergoing endoscopic surgery for arterial bypass.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

3 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K101274, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Jun 2, 2026Z-2741-2026Getinge identified that Vasoview Hemopro 2 devices from Lot 3000547797 were exposed to an out-of-specification sterilization dose during a gamma sterilization run (Run ID 1183-28782A) performed by STERIS at its…Open
Mar 20, 2025Z-1622-2025There have been at least 12 complaints from users as of March 3, 2025, reporting that they identified the ceramic C-Ring to be broken in half during use of the device. The ceramic C-Ring has a potential to break while…Open
Mar 10, 2025Z-1577-2025Incorrect resistor utilized in the VH-3010 Power Supply, which may cause the jaws of the HemoPro 2 to not adequately heat up.Open
Dec 9, 2024Z-0712-2025, Z-0711-2025Potential for two issues: 1. Bent or detached heater wire; 2. Silicone peeling or detaching from the Jaws of the Harvesting ToolOpen
May 31, 2024Z-2338-2024There were 27 complaints between March 22, 2024, and April 30, 2024, related to VH-3500 and VH-4000, reporting that the C-Ring wire was straightened (due to insufficient bend radius) and that the distance between C-Ring…Open
May 17, 2024Z-2175-2024Potential for there to be a fluid ingress into the handle of the harvesting tool, which may lead to the device not providing cautery or providing cautery when not intended.Open
Mar 6, 2023Z-1336-2023Some batches of product were not sterilized to their minimum sterilization specification.Open
Dec 22, 2022Z-0995-2023Ink on the Vasoshield syringe Maquet logo may chip resulting in an unreasonable risk of harm to the patient. Potential harms may include foreign body reaction, coronary embolic event, and peripheral embolic event.Open
Oct 30, 2013Z-0466-2014Maquet has received an increase in the number of complaints for "self-activated/remains activated/overheats" for the Hemopro VH-30000, "Hemopro 1"Terminated

Other GEI clearances

The newest 510(k)s for the same product code; their median review was 167 days; see every GEI device

Record

Type
Traditional, third-party review
Received
May 6, 2010
Review panel
General, Plastic Surgery
Regulation
21 CFR 878.4400

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files