Closed Fusion Plate, Mtp Plate, Talus Plate, Model 824077021, 22 and 1 more
Plate, Fixation, Bone. 510(k) K101240, from DePuy Orthopaedics, Inc. (Warsaw, IN). Product code HRS, class 2.
As of Nov 17, 2026, K101240 is an FDA 510(k) held by DePuy Orthopaedics, Inc., substantially equivalent on Jul 23, 2010 after 80 days, with no recalls tied to it in FDA's recall database.
- Substantially equivalent: Jul 23, 2010
- Review time: 80 days
- Recalls: 0
- Cited as a predicate by: 9 devices
Indications for Use
As the 510(k) summary states them, read from FDA's summary PDFThe DePuy ALPS Small Bone Locked Plating System is intended for use in stabilization and fixation of fractures, revision procedures, fusions, reconstructions (osteotomy) and non-unions of the bones of the hand, foot, wrist, ankle, finger, toe, humerus, olecranon, clavicle, scapula and pelvis, particularly in osteopenic bone.
Predicate Devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- Lateral Column Length Plate, Model 824073000-3, Locked Fusion Plate, Model 824077010,…K091294 · DePuy Orthopaedics, Inc. · Jul 24, 2009
- Fragment Plate SystemK061748 · Hand Innovations, LLC · Jul 17, 2006
Cited as a Predicate
9 later 510(k) summaries name this device as a predicate- Treace Medical Concepts (TMC) Plating SystemK183321 · Treace Medical Concepts, Inc. · Jan 25, 2019
- VariAx 2 SystemK173135 · Stryker GmbH · Nov 27, 2017
- Life Spine Foot and Ankle Plating SystemK172973 · Life Spine, Inc. · Dec 21, 2017
- I.T.S. Hand Locking Plates System - HLSK142418 · I.T.S. GmbH · Sep 26, 2014
- Variax 2 SystemK140376 · Stryker Trauma AG · Apr 3, 2014
- Lapidus Plate, 3.5MM Inline Fusion Plate, 3.5MM Mtp Plate (Left And Right)K131670 · Biomet, Inc. · Jul 3, 2013
- Checkmate Metatarso-Phalangeal (Mtp) Arthrodesis SystemK113762 · Arthrosurface, Inc. · Feb 6, 2012
- Claw Ii Polyaxial Compression System And Ortholoc 3DSI Locking ScrewsK113014 · Wrightmedicaltechnologyinc · Jan 31, 2012
- Trauma Internal Fixation SystemsK111663 · DePuy Orthopaedics, Inc. · Mar 12, 2012
Other HRS Clearances
The newest 510(k)s for the same product code; their median review was 79 days; see every HRS device- Arthrex Plates, Screws and Pins, and WashersK263602 · Arthrex, Inc. · Oct 29, 2026
- Trochlear Salvage Plating SystemK261135 · Skeletal Dynamics, Inc. · Oct 26, 2026
- ANTHEM Fracture SystemK262779 · Globus Medical, Inc. · Oct 23, 2026
- Tempo Plating SystemK263407 · Pace Surgical · Oct 23, 2026
- Park's Pectus SystemK263269 · Tdm Co., Ltd. · Oct 15, 2026
- TDM Large Bone Plate and Screw SystemK261611 · Tdm Co., Ltd. · Oct 2, 2026
- Alians ForearmK262649 · Newclip Technics · Sep 28, 2026
- ProRib Fixation SystemK261892 · Costa Surgical, Inc. · Sep 23, 2026
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- May 4, 2010
- Review panel
- Orthopedic
- Regulation
- 21 CFR 888.3030
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files