Arca
Skip to content
Tools
Tools

Iolmaster 500

Biomicroscope, Slit-Lamp, Ac-Powered. 510(k) K101182, from Carl Zeiss Meditec, AG (Dublin, CA). Product code HJO, class 2.

As of Nov 20, 2026, K101182 is an FDA 510(k) held by Carl Zeiss Meditec, AG, substantially equivalent on Nov 3, 2010 after 190 days, with 4 recalls tied to it in FDA's recall database.

  • Review time: 190 days
  • Recalls: 4
  • Cited as a predicate by: 1 device

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The IOLMaster is intended for the biometric determination of ocular measurements for axial length, anterior chamber depth, corneal radius, white-to-white (WTW), and for the measurement of pupil size and deviation of the visual axis from the center of the pupil.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

1 later 510(k) summary names this device as a predicate
  • Iolmaster 500K122418 · Carl Zeiss Meditec, AG · Apr 12, 2013

Recalls

Recalls FDA ties to K101182, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Oct 27, 2015Z-0357-2016, Z-0358-2016IOL Master software versions 7.5 and 7.7 calculation printouts and exported reports can contain the wrong IOL power data.Terminated
Oct 23, 2013Z-0333-2014, Z-0334-2014The user database delivered for download contained an erroneous value. Specifically the "aO" constant of the lens AMO Sensar AR40E is listed as "0.0" in the internal lens database of the IOLMaster. The correct value is…Terminated

Other HJO clearances

The newest 510(k)s for the same product code; their median review was 108 days; see every HJO device

Record

Type
Traditional
Received
Apr 27, 2010
Review panel
Ophthalmic
Regulation
21 CFR 886.1850

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Esc

Type a name: a tool, a company, a drug or brand, a condition, a gene or a state. Identifiers like K231234, NCT05012345 or NDA 215866 open their record.

Logos are each company's own site icon, matched by name through Wikidata, or logos provided by Logo.dev.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.