Arca
Skip to content
Tools
Tools

Spinesmith Cynch Spinal System

Intervertebral Fusion Device With Bone Graft, Lumbar. 510(k) K101085, from Spine Smith Partners L.P. (Austin, TX). Product code MAX, class 2.

As of Nov 19, 2026, K101085 is an FDA 510(k) held by Spine Smith Partners L.P., substantially equivalent on Jul 1, 2010 after 73 days, with 1 recall tied to it in FDA's recall database.

  • Review time: 73 days
  • Recalls: 1

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The Cynch System is indicated intervertebral body fusion of the lumbar spine, from L2 to S1, in skeletally mature patients who have had six months of non-operative treatment. The device is intended for use at either one level or two contiguous levels for the treatment of degenerative disc disease (DDD) with up to Grade I spondylolisthesis.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Recalls

Recalls FDA ties to K101085, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
StartedRecallStatus
Aug 24, 2012Z-0932-2013Implants were incorrectly laser marked as 23mm instead of 28mm.Terminated

Other MAX clearances

The newest 510(k)s for the same product code; their median review was 142 days; see every MAX device

Record

Type
Special
Received
Apr 19, 2010
Review panel
Orthopedic
Regulation
21 CFR 888.3080

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Esc

Type a name: a tool, a company, a drug or brand, a condition, a gene or a state. Identifiers like K231234, NCT05012345 or NDA 215866 open their record.

Logos are each company's own site icon, matched by name through Wikidata, or logos provided by Logo.dev.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.