Modification To Synthes Matrix System
Thoracolumbosacral Pedicle Screw System. 510(k) K100952, from Synthes Spine (West Chester, PA). Product code NKB, class 2.
As of Nov 17, 2026, K100952 is an FDA 510(k) held by Synthes Spine, substantially equivalent on Aug 5, 2010 after 121 days, with 2 recalls tied to it in FDA's recall database.
- Substantially equivalent: Aug 5, 2010
- Review time: 121 days
- Recalls: 2
- Cited as a predicate by: 10 devices
Indications for Use
As the 510(k) summary states them, read from FDA's summary PDFAUG 05 2010 510(k) Number: K100952 Device Name: Synthes Matrix System The Synthes USS are non-cervical spinal fixation devices intended for posterior pedicle screw fixation (T1-S2/ilium), posterior hook fixation (T1-L5), or anterolateral fixation (T8-L5).
Predicate Devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- Synthes Matrix SystemK092929 · Synthes (Usa) · Dec 29, 2009
- Synthes Matrix SystemK100634 · Synthes (Usa) · Jul 1, 2010
- Synthes (U.S.A.) Universal Sacral SystemK963045 · Synthes (Usa) · Feb 13, 1997
- Uss Iliosacral, Uss PolyaxialK082572 · Synthes Spine · Nov 24, 2008
- Click'XK992739 · Synthes Spine · Feb 8, 2000
- Synthes Pangea SystemK052123 · Synthes (Usa) · Sep 23, 2005
- Uss VasK002517 · Synthes (Usa) · Sep 12, 2000
Cited as a Predicate
10 later 510(k) summaries name this device as a predicate- OLYMPIC Posterior Spinal Fixation SystemK181139 · Astura Medical, LLC · Jul 6, 2018
- OLYMPIC Posterior Spinal Fixation SystemK153446 · Astura Medical · Apr 8, 2016
- M.U.S.T Pedicle Screw SystemK141988 · Medacta International · Aug 19, 2014
- Creo Stabilization SystemK124058 · Globus Medical, Inc. · May 13, 2013
- Easyspine SystemK123134 · Ldr Spine USA, Inc. · Feb 1, 2013
- Medyssey Cannulated Pedicle ScrewK121670 · Medyssey Co, Ltd. · Jan 25, 2013
- M.U.S.T Pedicle Screw SystemK121115 · Medacta International · Jul 18, 2012
- Synthes Uss ConnectorK120571 · Synthes Spine · May 14, 2012
- Synthes Uss ConnectorsK113149 · Synthes Spine · Feb 1, 2012
- Synthes UssK111358 · Synthes Spine · Nov 18, 2011
Recalls
Recalls FDA ties to K100952, newest first; each opens FDA's recall record| Oct 23, 2014 | Z-0226-2015Two nonconforming parts of the Titanium Polyaxial Reduction Head for Titanium Matrix Spine Screws may have been inadvertently packaged and shipped. It is believed that the inner diameter of the Reduction Head may not… | Terminated |
| Mar 11, 2011 | Z-3207-2011Labeling Correction -- The locking cap was not able to be loosened, resulting in the breakage of the screwdriver or the rod to be cut and the assembly to be removed (screw, cap, and rod). This poses a potential risk of… | Terminated |
Other NKB Clearances
The newest 510(k)s for the same product code; their median review was 90 days; see every NKB device- Virata Spinal Fixation SystemK262731 · SeaSpine Orthopedics Corporation · Oct 29, 2026
- AnyPlus II Spinal Fixation SystemK261217 · GS Medical Co., Ltd. · Oct 2, 2026
- NOVIX Pedicle Screw SystemK263062 · Tdm Co., Ltd. · Oct 1, 2026
- Duet Spinal Fixation SystemK262712 · Duet Spine Holdings, LLC · Sep 30, 2026
- SAS MIS Pelvic Incidence RodsK263172 · Spinal Alignment Solutions, Inc. · Sep 30, 2026
- Kentrospine Viktr-MIS® Pedicle screw systemK252139 · Rivarp Medical Pvt., Ltd. · Sep 27, 2026
- iFuse Granite Onyx Implant SystemK262067 · SI-BONE, Inc. · Sep 23, 2026
- FRAXION Spinal Fixation SystemK262091 · Ortler Global Tibbi Gerecler Anonim Sirketi · Sep 17, 2026
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Special
- Received
- Apr 6, 2010
- Review panel
- Orthopedic
- Regulation
- 21 CFR 888.3070
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files