Cobas 8000 Modular Series Analyzer

Albumin, Antigen, Antiserum, Control. 510(k) K100853, from Roche Diagnostics Corp. (Indianapolis, IN). Product code DCF, class 2.

As of Nov 17, 2026, K100853 is an FDA 510(k) held by Roche Diagnostics Corp., substantially equivalent on Sep 9, 2010 after 167 days, with 11 recalls tied to it in FDA's recall database.

  • Substantially equivalent: Sep 9, 2010
  • Review time: 167 days
  • Recalls: 11
  • Cited as a predicate by: 9 devices

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

The cobas 8000 Modular Analyzer Series is a fully automated system for clinical chemistry analysis, intended for the in vitro qualitative and quantitative determination of analytes in body fluids. It is optimized for high throughput workloads in the professional environment using a combination of ion selective electrodes (ISE) and a photometric analysis unit.

Predicate Devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a Predicate

9 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K100853, newest first; each opens FDA's recall record
Dec 8, 2020Z-0754-2021Potential for Changed Configuration Settings on the cobas 8000 modular analyzer series/cobas pro integrated solutions, may cause incorrect results in several affected parameters. In the case of poor sample quality,…Open, Classified
Jul 18, 2019Z-2477-2019Quality issue with high pressure solenoid valvesTerminated
Mar 15, 2019Z-1249-2019The firm has confirmed complaints regarding signal drops on the cobas e 801 module after a ProCell II M bottle changeover, which may lead to incorrect medical decisions with respect to diagnostics and patient treatment.Terminated
Apr 3, 2018Z-2136-2018The syringe plunger on the device may be mounted in a tilted position, which applies mechanical stress to the plunger. In the worst case scenario, the plunger may crack leading to impaired syringe function.Terminated
Jul 28, 2017Z-0675-2018A software failure may incorrectly set the system settings to "default" settings, creating a risk of incorrect results.Terminated
Jul 18, 2017Z-0910-2018A warning is being added to the cobas 6000 analyzer and the cobas 8000 modular analyzer series Operators Manual to prevent operator injury during maintenance.Terminated
Jun 29, 2017Z-0378-2018Roche has confirmed that a possible sample mismatch issue may occur on the MODULAR ANALYTICS E 170 module, cobas e 601 analyzer, or cobas e 602 analyzer due to a software limitation.Terminated
Mar 21, 2017Z-2049-2017Due to a design change in 2015, corrosion on the sample probe electrical contacts can occur and lead to a disturbance of the sample liquid level detection (LLD). When disturbance of the sample liquid level detection…Terminated
Jul 15, 2016Z-0097-2017Roche has received four reports of injury caused by the sharp edge of the center guide rail of an analyzers sample rack tray. The cleaning procedure for the sample rack tray has been revised to clarify the cleaning…Terminated
Mar 13, 2014Z-1449-2014The Initial Cassette Volume Check (ICVC) is not activated on the cobas c 502 module of the cobas 8000 modular analyzer series when manually filled cobas c packs are loaded. This issue occurs with tests that use cobas c…Terminated
Apr 4, 2013Z-1368-2013When using the cobas e 602 module, Diluent Multiassay may be used for autodilution of certain assays instead of Diluent Universal. This may cause incorrect results under specific circumstances if all of the following…Terminated

Record

Type
Special
Received
Mar 26, 2010
Review panel
Immunology
Regulation
21 CFR 866.5040

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Logos are each company's own site icon, matched by name through Wikidata.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.