Arca
Skip to content
Tools
Tools

Raystation Version 1.0

System, Planning, Radiation Therapy Treatment. 510(k) K100552, from Raysearch Laboratories AB (Stockholm, SE). Product code MUJ, class 2.

As of Nov 20, 2026, K100552 is an FDA 510(k) held by Raysearch Laboratories AB, substantially equivalent on Mar 12, 2010 after 14 days, with 4 recalls tied to it in FDA's recall database.

  • Review time: 14 days
  • Recalls: 4
  • Cited as a predicate by: 2 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

RayStation is a software system designed for treatment planning and analysis of radiation therapy. The treatment plans provide treatment unit set-up parameters and estimates of dose distributions expected during the proposed treatment, and may be used to administer treatments after review and approval by the intended user. The system functionality can be configured based on user needs.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

2 later 510(k) summaries name this device as a predicate
  • RaystationK120387 · Raysearch Laboratories AB · Apr 6, 2012
  • RadianceK112060 · Gmv Aerospace and Defence S.A. · Jan 6, 2012

Recalls

Recalls FDA ties to K100552, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Feb 14, 2014Z-1481-2014RaySearch Laboratories has recalled "RaySearch Ray Station 2.0, 2.5, 3.0, 3.5 and 4.0" software due to dose miscalculations for bolus/external/fixation support structures on CT-image.Terminated
Feb 6, 2014Z-1307-2014Potential for dose errors due to software program errors.Terminated
Dec 6, 2011Z-1317-2012An issue was found with the dose calculation in the RayStation 2.0, that could result in a potential overestimation of the dose. This could lead to the potential under-dosage of a patient.Terminated
Oct 19, 2011Z-0359-2012The firm has found an issue with DICOM RT Image representation of DRR images from RayStation 2.0 which needs to be clarified in the user manual. The symptoms involve the Pixel Spacing and Image Position attributes in…Terminated

Other MUJ clearances

The newest 510(k)s for the same product code; their median review was 237 days; see every MUJ device

Record

Type
Traditional, third-party review
Received
Feb 26, 2010
Review panel
Radiology
Regulation
21 CFR 892.5050

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files