Arca
Skip to content
Tools
Tools

Utaswako I30 (Includes Accessories), Utaswako AFP-L3, Calibrator Set, Control L, Control H, Sample Dilution

Test, Alpha Fetoprotein L3 Subfraction (Afp-L3%), For Hepatocellular Carcinoma Risk Assessment. 510(k) K100464, from Wako Chemicals USA, Inc. (Mountain View, CA). Product code NSF, class 2.

As of Nov 25, 2026, K100464 is an FDA 510(k) held by Wako Chemicals USA, Inc., substantially equivalent on Feb 23, 2011 after 370 days, with 2 recalls tied to it in FDA's recall database.

  • Review time: 370 days
  • Recalls: 2

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

___ K100464 Device Name: Wako μTASWako DCP Calibrator Set, Wako μTASWako DCP Control L, and Wako μTASWako DCP Control H The Wako μTASWako DCP Calibrator Set is designed to be used with the Wako μTASWako DCP Immunological Test System for the quantitative determination of DCP in human serum.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF
  • K041847K041847

Recalls

Recalls FDA ties to K100464, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Feb 18, 2021Z-1412-2021Potentially defective Chip Cassette units, when used for DCP or AFP-L3% measurement, incorrect DCP and AFP-L3% values may resultTerminated
Apr 30, 2013Z-1831-2013Product fails to meet the accuracy at the low end of the measuring range.Terminated

Record

Type
Traditional
Received
Feb 18, 2010
Review panel
Immunology
Regulation
21 CFR 866.6030

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files