Miar (Minimally Invasive Aortic Root) Cannula With Flow-Guard, Models 11012L and 11014L

Catheter, Cannula And Tubing, Vascular, Cardiopulmonary Bypass. 510(k) K100274, from Medtronic, Inc. (Minneapolis, MN). Product code DWF, class 2.

As of Nov 27, 2026, K100274 is an FDA 510(k) held by Medtronic, Inc., substantially equivalent on May 10, 2010 after 101 days, with 1 recall tied to it in FDA's recall database.

  • Review time: 101 days
  • Recalls: 1

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The MiAR™ cannula is intended for use during cardiopulmonary bypass for the delivery of cardioplegia for up to 6 hours. The cannula may also be used to aspirate air from the aorta at the conclusion of the bypass procedure. It is indicated for use during cardiac surgery for median sternotomy or minimally invasive (mini-sternotomy or right thoracotomy) access using direct visualization techniques.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Recalls

Recalls FDA ties to K100274, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
StartedRecallStatus
Feb 5, 2025Z-1306-2025Unexpected loose material in the male luer used in the aortic root cannula has been identified. Potential harms when identified prior to use - procedure delay. Potential harm if not identified and used - stroke…Open

Other DWF clearances

The newest 510(k)s for the same product code; their median review was 91 days; see every DWF device

Record

Type
Traditional
Received
Jan 29, 2010
Review panel
Cardiovascular
Regulation
21 CFR 870.4210

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files