Infinity Acute Care System Workstation Neonatal Care

Ventilator, Continuous, Facility Use. 510(k) K093632, from Drager Medical AG & Co. Kgaa (Telford, PA). Product code CBK, class 2.

As of Nov 17, 2026, K093632 is an FDA 510(k) held by Drager Medical AG & Co. Kgaa, substantially equivalent on Apr 5, 2010 after 133 days, with 6 recalls tied to it in FDA's recall database.

  • Substantially equivalent: Apr 5, 2010
  • Review time: 133 days
  • Recalls: 6
  • Cited as a predicate by: 2 devices

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

The Infinity Acute Care Systems Workstations Neonatal Care consist of monitoring and control displays and additional therapy units. They are intended to be used as integrated, networked, and configurable workstations to provide neonatal care specific therapy. The Infinity Acute Care Systems Workstations Neonatal Care are intended to be used by qualified and trained medical personnel.

Predicate Devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a Predicate

2 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K093632, newest first; each opens FDA's recall record
Feb 14, 2022Z-0871-2022Cybersecurity Improvement Action-not equipped against potential cyber security threatsTerminated
Sep 10, 2021Z-0109-2022Software 2.51 and Lower with Installed CO2 Measurement Option, may result in Restart of the Ventilator, brief cessation of ventilation can occur and loss of PEEP for approx. 8 secondsOpen, Classified
Feb 4, 2016Z-0827-2016The battery capacity of the optional PS500 power supply unit for the Infinity (ACS) Workstation Critical and Neonatal Care was rapidly and unexpectedly reduced, despite using the most recent power supply firmware…Terminated
Dec 1, 2015Z-0436-2016The firm became aware of cases in which the battery run times of the optional PS 500 power supply unit with the Infinity Workstation Critical Care (Evita Infinity V 500) were unexpectedly short due to the design of the…Terminated
Jul 13, 2015Z-2343-2015The battery capacity of the optional PS500 of an IACS Workstation Critical Care (Evita Infinity V500) did not last as long as expected.Terminated
Feb 7, 2014Z-1298-2014The battery capacity of the optional PS500 Power Supply Unit of the Infinity ACS Workstation Critical Care (Evita Infinity V500) did not last as long as expected.Terminated

Other CBK Clearances

The newest 510(k)s for the same product code; their median review was 235 days; see every CBK device

Record

Type
Traditional
Received
Nov 23, 2009
Review panel
Anesthesiology
Regulation
21 CFR 868.5895

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Logos are each company's own site icon, matched by name through Wikidata.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.