Liaison N-Tact Pth Calibration Verifiers Model 310913
Single (Specified) Analyte Controls (Assayed And Unassayed). 510(k) K093498, from DiaSorin, Inc. (Stillwater, MN). Product code JJX, class 1.
As of Nov 20, 2026, K093498 is an FDA 510(k) held by DiaSorin, Inc., substantially equivalent on Dec 14, 2009 after 32 days, with no recalls tied to it in FDA's recall database.
- Review time: 32 days
- Recalls: 0
- Cited as a predicate by: 2 devices
- Liaison N-Tact PthK033426 · DiaSorin, Inc. · Feb 12, 2004
- LIASON 1-84 PTH Assay, LIASON 1-84 PTH Control Set, LIASON 1-84 PTH Calibration VerifiersK150879 · DiaSorin, Inc. · Jun 22, 2015
- Liaison N-Tact Pth Gen Ii, Control Set, Calibration VerifiersK132515 · DiaSorin, Inc. · Nov 8, 2013
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Nov 12, 2009
- Review panel
- Clinical Chemistry
- Regulation
- 21 CFR 862.1660
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files