Depuy Fracture And Fusion Plating System
Plate, Fixation, Bone. 510(k) K093474, from DePuy Orthopaedics, Inc. (Warsaw, IN). Product code HRS, class 2.
As of Nov 17, 2026, K093474 is an FDA 510(k) held by DePuy Orthopaedics, Inc., substantially equivalent on Feb 2, 2010 after 88 days, with 2 recalls tied to it in FDA's recall database.
- Substantially equivalent: Feb 2, 2010
- Review time: 88 days
- Recalls: 2
- Cited as a predicate by: 4 devices
Indications for Use
As the 510(k) summary states them, read from FDA's summary PDFThe DePuy Fracture and Fusion Plating System is intended for use in stabilization and fixation of fractures, revision procedures, fusions, reconstructions (osteotomy) and non-unions of the bones of the hand, foot, wrist, ankle, finger, toe, humerus, olecranon, clavicle, scapula and pelvis, particularly in osteopenic bone.
Cited as a Predicate
4 later 510(k) summaries name this device as a predicate- OrthoPediatrics Wrist Fusion PlateK172425 · Orthopediatrics, Inc. · Dec 6, 2017
- A.L.P.S. Calcaneal Plating SystemK132898 · Biomet Manufacturing Corp · Nov 19, 2013
- 2.5MM Inline Fusion PlateK131867 · Biomet, Inc. · Oct 29, 2013
- Trauma Internal Fixation SystemsK111663 · DePuy Orthopaedics, Inc. · Mar 12, 2012
Recalls
Recalls FDA ties to K093474, newest first; each opens FDA's recall record| Oct 25, 2017 | Z-1306-2018The drill guides were manufactured with the incorrect color anodization. | Terminated |
| Jun 4, 2015 | Z-2046-2015The thread location on one of the thread holes is offset from the axis of the predrill hole. The threads are too deep on one side and too shallow on the other. The locking screw may back out of the plate if it does not… | Terminated |
Other HRS Clearances
The newest 510(k)s for the same product code; their median review was 79 days; see every HRS device- Arthrex Plates, Screws and Pins, and WashersK263602 · Arthrex, Inc. · Oct 29, 2026
- Trochlear Salvage Plating SystemK261135 · Skeletal Dynamics, Inc. · Oct 26, 2026
- ANTHEM Fracture SystemK262779 · Globus Medical, Inc. · Oct 23, 2026
- Tempo Plating SystemK263407 · Pace Surgical · Oct 23, 2026
- Park's Pectus SystemK263269 · Tdm Co., Ltd. · Oct 15, 2026
- TDM Large Bone Plate and Screw SystemK261611 · Tdm Co., Ltd. · Oct 2, 2026
- Alians ForearmK262649 · Newclip Technics · Sep 28, 2026
- ProRib Fixation SystemK261892 · Costa Surgical, Inc. · Sep 23, 2026
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Nov 6, 2009
- Review panel
- Orthopedic
- Regulation
- 21 CFR 888.3030
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files