Arca
Skip to content

Depuy Fracture And Fusion Plating System

Plate, Fixation, Bone. 510(k) K093474, from DePuy Orthopaedics, Inc. (Warsaw, IN). Product code HRS, class 2.

As of Nov 17, 2026, K093474 is an FDA 510(k) held by DePuy Orthopaedics, Inc., substantially equivalent on Feb 2, 2010 after 88 days, with 2 recalls tied to it in FDA's recall database.

  • Substantially equivalent: Feb 2, 2010
  • Review time: 88 days
  • Recalls: 2
  • Cited as a predicate by: 4 devices

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

The DePuy Fracture and Fusion Plating System is intended for use in stabilization and fixation of fractures, revision procedures, fusions, reconstructions (osteotomy) and non-unions of the bones of the hand, foot, wrist, ankle, finger, toe, humerus, olecranon, clavicle, scapula and pelvis, particularly in osteopenic bone.

Cited as a Predicate

4 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K093474, newest first; each opens FDA's recall record
Oct 25, 2017Z-1306-2018The drill guides were manufactured with the incorrect color anodization.Terminated
Jun 4, 2015Z-2046-2015The thread location on one of the thread holes is offset from the axis of the predrill hole. The threads are too deep on one side and too shallow on the other. The locking screw may back out of the plate if it does not…Terminated

Other HRS Clearances

The newest 510(k)s for the same product code; their median review was 79 days; see every HRS device

Record

Type
Traditional
Received
Nov 6, 2009
Review panel
Orthopedic
Regulation
21 CFR 888.3030

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files