Arca
Skip to content

Vanguard 360 Revision Knee System

Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer. 510(k) K093293, from Biomet, Inc. (Warsaw, IN). Product code JWH, class 2.

As of Nov 17, 2026, K093293 is an FDA 510(k) held by Biomet, Inc., substantially equivalent on Oct 22, 2010 after 366 days, with 5 recalls tied to it in FDA's recall database.

  • Substantially equivalent: Oct 22, 2010
  • Review time: 366 days
  • Recalls: 5
  • Cited as a predicate by: 6 devices

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

The indications for the Vanguard™ 360 Revision Knee System are as follows: 1. Painful and disabled knee joint resulting from osteoarthritis, rheumatoid arthritis, traumatic arthritis where one or more compartments are involved. 2. Correction of varus, valgus or posttraumatic deformity. 3.

Predicate Devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a Predicate

6 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K093293, newest first; each opens FDA's recall record
Dec 1, 2020Z-0723-2021Incorrect bolt -tibial bolts in package instead of femoral bolts may delay procedureTerminated
Aug 2, 2018Z-3250-2018The implant only has one bolt in the sterile package but it should have had two.Terminated
Aug 2, 2018Z-3249-2018The implant only has one bolt in the sterile package but it should have had two.Terminated
Aug 8, 2016Z-0003-2017The drill bit used to drill the hex pre-drilled hole in the screw head of the Tibial Augment Bolts included in the Vanguard 360 Revision Knee System was larger than allowable per the process specification. As a result,…Terminated
Jul 1, 2014Z-2242-2014The Vanguard 360 Femoral Boss Reamers were manufactured oversized to the print. The instrument may not fit into the mating component.Terminated

Other JWH Clearances

The newest 510(k)s for the same product code; their median review was 88 days; see every JWH device

Record

Type
Traditional
Received
Oct 21, 2009
Review panel
Orthopedic
Regulation
21 CFR 888.3560

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Logos are each company's own site icon, matched by name through Wikidata.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.