Arca
Skip to content
Tools
Tools

5 MM Harmonic Ace Instrument (Used With Da Vinci IS1200 & IS2000/IS3000 System)

System, Surgical, Computer Controlled Instrument. 510(k) K093217, from Intuitive Surgical, Inc. (Sunnyvale, CA). Product code NAY, class 2.

As of Nov 20, 2026, K093217 is an FDA 510(k) held by Intuitive Surgical, Inc., substantially equivalent on Jan 21, 2010 after 100 days, with 3 recalls tied to it in FDA's recall database.

  • Review time: 100 days
  • Recalls: 3
  • Cited as a predicate by: 2 devices

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

2 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K093217, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Jan 16, 2019Z-0841-2019During an internal inspection, Intuitive found that the sterile packaging of the da Vinci S/Si Harmonic ACE Inserts may become damaged during transit. The potential damage includes pinholes and /or small openings (less…Open
Jan 21, 2015Z-1118-2015, Z-1117-2015Internal labeling review found that the IFU for the Harmonic Ace incorrectly instructs the user on how to manually open the grips.Terminated

Other NAY clearances

The newest 510(k)s for the same product code; their median review was 89 days; see every NAY device

Record

Type
Traditional
Received
Oct 13, 2009
Review panel
Gastroenterology, Urology
Regulation
21 CFR 876.1500

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files