S-Lif Intervertebral Body Fusion Device
Intervertebral Fusion Device With Bone Graft, Lumbar. 510(k) K092815, from Spinefrontier, Inc. (Beverly, MA). Product code MAX, class 2.
As of Nov 22, 2026, K092815 is an FDA 510(k) held by Spinefrontier, Inc., substantially equivalent on Jun 30, 2010 after 289 days, with 6 recalls tied to it in FDA's recall database.
- Review time: 289 days
- Recalls: 6
- Cited as a predicate by: 1 device
The S-LIF™ Intervertebral Body Fusion Device is a spinal intervertebral body fusion device intended for intervertebral body fusion of the spine of skeletally mature patients, using autogenous bone graft to facilitate fusion. The device is indicated for use in patients with degenerative disc disease (DDD) at one or two contiguous spinal levels from L2-S1.
- Spinefrontier Dorado VbrK072289 · Spinefrontier, Inc. · Nov 1, 2007
- NuVasive Coroent SystemK071795 · Nuvasive, Inc. · Dec 4, 2007
- Oracle Spacer, Opal SpacerK072791 · Synthes Spine · Dec 26, 2007
- Spinefrontier Lumbar Ibf SystemK111553 · Spinefrontier, Inc. · Sep 28, 2011
| Aug 28, 2013 | Z-0274-2014Fit variation among all lots of S-LIFT extension shims with corresponding retractor blades. The firm's investigation revealed that extension shims may not fit down all retractor blades. | Terminated |
| May 24, 2013 | Z-1665-2013The outer sleeve would not fit over the S-LIFT Inserter assembly because the hole at the top of the sleeve handle was too small. | Terminated |
| May 17, 2013 | Z-1664-2013The S-LIFT Instrument Case contained a bracket with peeling nylon coating. | Terminated |
| Aug 29, 2012 | Z-1932-2013During an operation the strike plate of a T-Handle loosened and separated from the T-Handle. | Terminated |
| Aug 22, 2012 | Z-0417-2013When using the optional S-LIFT (also known as S-LIF) Implant Guides without stops, there is a known risk of moving the Implant Guides, and possibly the implant, too far into the disc space creating the potential for… | Terminated |
| Dec 23, 2011 | Z-1565-2014It has been verified through field use that there are two possible breakages that may occur when using SI50001 Revision D Lock Shaft Inserter, possibly resulting in instrument fragments entering the sterile field:… | Terminated |
- 4CIS Blue Dolphin Cage SystemK252865 · Solco Biomedical Company India Private Limited · Nov 6, 2026
- Symmetrix™ PX Spinal SystemK262572 · Medtronic Sofamor Danek USA, Inc. · Oct 29, 2026
- Calibrate MAX System; Calibrate NanoTec MAX System; Calibrate CCX Interbody System; Calibrate NanoTec CCX Interbody…K262646 · Alphatec Spine, Inc. · Oct 21, 2026
- Scala® TLK261049 · Spineart SA · Aug 20, 2026
- TruLift Expandable SystemK261430 · Life Spine, Inc. · Jul 28, 2026
- Catalyft PL Expandable Interbody SystemK253011 · Medtronic Sofamor Danek USA, Inc. · Jul 17, 2026
- Symmetrix™ L Spinal SystemK261020 · Medtronic Sofamor Danek USA, Inc. · Jul 14, 2026
- Admiral ACP System; Explorer TO System; Laminoplasty System; Manta Ray TDF Spacer; Meridian Interbody System; Meridian…K253420 · SeaSpine Orthopedics Corporation · Jul 1, 2026
- Document
- 510(k) summary (PDF) ↗
- Type
- Abbreviated
- Received
- Sep 14, 2009
- Review panel
- Orthopedic
- Regulation
- 21 CFR 888.3080
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files