Etrax Needle System, Etrax Variable Angle Needle Guidance System, Virtutrax Universal Tracker

System, Imaging, Pulsed Echo, Ultrasonic. 510(k) K092619, from Civco (Kalona, IA). Product code IYO, class 2.

As of Nov 18, 2026, K092619 is an FDA 510(k) held by Civco, substantially equivalent on Nov 13, 2009 after 79 days, with 17 recalls tied to it in FDA's recall database.

  • Substantially equivalent: Nov 13, 2009
  • Review time: 79 days
  • Recalls: 17
  • Cited as a predicate by: 5 devices

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

The device is intended to provide physicians with tools for electromagnetic tracking of instruments with respect to image data.

Predicate Devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a Predicate

5 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K092619, newest first; each opens FDA's recall record
Mar 2, 2026Z-1738-2026There was an error in inspection and programming of the eTRAX needle sensor for Aurora trackers. The result is a potential for the needle tip position to be incorrectly identified on the user interface.Open, Classified
Mar 2, 2026Z-1734-2026There was an error in inspection and programming of the eTRAX needle sensor for Aurora trackers. The result is a potential for the needle tip position to be incorrectly identified on the user interface.Open, Classified
Mar 2, 2026Z-1733-2026There was an error in inspection and programming of the eTRAX needle sensor for Aurora trackers. The result is a potential for the needle tip position to be incorrectly identified on the user interface.Open, Classified
Mar 2, 2026Z-1735-2026There was an error in inspection and programming of the eTRAX needle sensor for Aurora trackers. The result is a potential for the needle tip position to be incorrectly identified on the user interface.Open, Classified
Mar 2, 2026Z-1732-2026There was an error in inspection and programming of the eTRAX needle sensor for Aurora trackers. The result is a potential for the needle tip position to be incorrectly identified on the user interface.Open, Classified
Mar 2, 2026Z-1737-2026There was an error in inspection and programming of the eTRAX needle sensor for Aurora trackers. The result is a potential for the needle tip position to be incorrectly identified on the user interface.Open, Classified
Mar 2, 2026Z-1731-2026There was an error in inspection and programming of the eTRAX needle sensor for Aurora trackers. The result is a potential for the needle tip position to be incorrectly identified on the user interface.Open, Classified
Mar 2, 2026Z-1736-2026There was an error in inspection and programming of the eTRAX needle sensor for Aurora trackers. The result is a potential for the needle tip position to be incorrectly identified on the user interface.Open, Classified
Oct 8, 2021Z-0395-2022There is potential that the VirtuTRAX device will not tighten onto the instrument to be used.Terminated
Oct 8, 2021Z-0398-2022There is potential that the VirtuTRAX device will not tighten onto the instrument to be used.Terminated
Oct 8, 2021Z-0397-2022There is potential that the VirtuTRAX device will not tighten onto the instrument to be used.Terminated
Oct 8, 2021Z-0396-2022There is potential that the VirtuTRAX device will not tighten onto the instrument to be used.Terminated
May 7, 2012Z-1952-2012The general purpose electromagnetic sensor was incorrectly calibrated causing potential variability in the projected position of the needle path with the ultrasound device.Terminated
May 7, 2012Z-1956-2012The general purpose electromagnetic sensor was incorrectly calibrated causing potential variability in the projected position of the needle path with the ultrasound device.Terminated
May 7, 2012Z-1954-2012The general purpose electromagnetic sensor was incorrectly calibrated causing potential variability in the projected position of the needle path with the ultrasound device.Terminated
May 7, 2012Z-1955-2012The general purpose electromagnetic sensor was incorrectly calibrated causing potential variability in the projected position of the needle path with the ultrasound device.Terminated
May 7, 2012Z-1953-2012The general purpose electromagnetic sensor was incorrectly calibrated causing potential variability in the projected position of the needle path with the ultrasound device.Terminated

Other IYO Clearances

The newest 510(k)s for the same product code; their median review was 234 days; see every IYO device

Record

Type
Traditional, third-party review
Received
Aug 26, 2009
Review panel
Radiology
Regulation
21 CFR 892.1560

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files