Arca
Skip to content
Tools
Tools

O-Arm Imaging System

Image-Intensified Fluoroscopic X-Ray System, Mobile. 510(k) K092564, from Medtronic Navigation, Inc. (Littleton) (Littleton, MA). Product code OXO, class 2.

As of Nov 18, 2026, K092564 is an FDA 510(k) held by Medtronic Navigation, Inc. (Littleton), substantially equivalent on Jun 17, 2010 after 301 days, with 18 recalls tied to it in FDA's recall database.

  • Review time: 301 days
  • Recalls: 18
  • Cited as a predicate by: 2 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The O-arm® Imaging System is a mobile x-ray system designed for 2D fluoroscopic and 3D imaging and is intended to be used where a physician benefits from 2D and 3D information of anatomic structures and objects with high x-ray attenuation such as bony anatomy and metallic objects. The O-arm® Imaging System is compatible with certain Image Guided Surgery Systems.

Cited as a predicate

2 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K092564, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
May 1, 2019Z-1932-2019The 0-arm 1000 Imaging System uses energy from the batteries to generate X-rays and move the system. If a charger board is not functioning properly, the batteries will not receive a full recharge of the potential…Terminated
Aug 30, 2018Z-0279-2019, Z-0271-2019, Z-0278-2019, Z-0280-2019, Z-0277-2019, Z-0274-2019, Z-0272-2019, Z-0276-2019, Z-0275-2019, Z-0273-2019Software anomalies were reported to affect the following system functionalities: startup and shutdown, system and network communication, motion control, image acquisition and output, dose reporting, and logs generation.…Terminated
Jan 25, 2018Z-0928-2018Software update ("SW Update 3.2.1") is being implemented to address the following issues: Inability to power the system after shutdown, System stays in standalone mode, Image reconstruction, System/Pendant bootup, Dose…Terminated
Mar 17, 2016Z-2082-2016It has been discovered that replacement x-ray generator and system motion batteries could become damaged during shipment, resulting in unacceptable risk of failure.Terminated
Jul 25, 2013Z-0973-2014Review of on-system electronic product labeling and Instructions for comply with applicable Use in response to a customer inquiry uncovered gaps in compliance FDA standards with the requirements for information to be…Terminated
Jan 31, 2013Z-0926-2013Potential failure of the braking system that controls the O-arm Imaging System gantry movement in the vertical directionTerminated
Jul 11, 2012Z-2244-2012A fault in a high voltage relay in the 0-arm Imaging System might cause the relay to overheat and burn.Terminated
Oct 15, 2010Z-0447-2011Safety Notice: Door Cable Periodic Maintenance Schedule and Cable Wear for the O-ARM Imaging System Breakage of the door cable during opening and closing may occur and cause physical harm or delay of therapyTerminated
Oct 12, 2010Z-1849-2012Software Version 3.1.1 does not meet navigational accuracy.Terminated

Other OXO clearances

The newest 510(k)s for the same product code; their median review was 152 days; see every OXO device

Record

Type
Abbreviated
Received
Aug 20, 2009
Review panel
Radiology
Regulation
21 CFR 892.1650

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files