Endogastric Solutions Esophyx2 System With Serosafuse Fastener and Accessories, Model 2.7.5
Endoscopic Suture/Plication System, Gastroesophageal Reflux Disease (Gerd). 510(k) K092400, from Endogastric Solutions, Inc. (Redmond, WA). Product code ODE, class 2.
As of Nov 19, 2026, K092400 is an FDA 510(k) held by Endogastric Solutions, Inc., substantially equivalent on Nov 6, 2009 after 92 days, with 2 recalls tied to it in FDA's recall database.
- Review time: 92 days
- Recalls: 2
- Cited as a predicate by: 1 device
The EndoGastric Solutions EsophyX2 Device with SerosaFuse Fastener and Accessories is indicated for use in transoral tissue approximation, full thickness plication and ligation in the GI tract and is indicated for the treatment of symptomatic chronic gastroesophageal reflux disease in patients who require and respond to pharmacological therapy.
- Egs Serosafuse Implantable Fasteners, Associated Delivery Devices and AccessoriesK071651 · Endogastric Solutions, Inc. · Sep 14, 2007
- Esophyx2 HDK142113 · Endogastric Solutions, Inc. · Sep 30, 2014
| Jun 6, 2013 | Z-1982-2013Endogastric Solutions, Inc. has received a limited number of reports relating to the loss of tissue mold control when operating the R2001 or R2002 EsophyX2 device. In one case, surgical intervention was necessary to… | Terminated |
| Feb 4, 2013 | Z-1136-2013EsophyX2Plus device was not validated under actual or simulated use conditions prior to distributing the product to physicians. | Terminated |
- GERDX-SystemK233240 · G-Surg GmbH · Jun 21, 2024
- EsophyX Z+ Device with SerosaFuse Fasteners and AccessoriesK240879 · Endogastric Solutions, Inc. · Apr 26, 2024
- Document
- 510(k) summary (PDF) ↗
- Type
- Special
- Received
- Aug 6, 2009
- Review panel
- Gastroenterology, Urology
- Regulation
- 21 CFR 876.1500
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files