Djo Surgical Revision Femoral Hip System, Model 428-14/24-140/200 and 1 more
Prosthesis, Hip, Semi-Constrained (Metal Uncemented Acetabular Component). 510(k) K092331, from Encore Medical L.P. (Austin, TX). Product code KWA, class 3.
As of Nov 18, 2026, K092331 is an FDA 510(k) held by Encore Medical L.P., substantially equivalent on Mar 3, 2010 after 211 days, with 1 recall tied to it in FDA's recall database.
- Substantially equivalent: Mar 3, 2010
- Review time: 211 days
- Recalls: 1
- Cited as a predicate by: 3 devices
Indications for Use
As the 510(k) summary states them, read from FDA's summary PDFModular Revision Hip Stem Indications for Use The Modular Revision Femoral Hip Stem is indicated for patients whose bone stock is of poor quality or inadequate for other reconstruction techniques as indicated by deficiencies of the femoral head, neck or portions of the proximal femur. It is intended for cementless revision hip arthroplasty on both uncemented and cemented femoral implants.
Predicate Devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- Zmr Hip System-Revision TaperK992667 · Zimmer, Inc. · Oct 27, 1999
- Zmr Hip System-Xl, Models 9921, 9922 And 9923K031572 · Zimmer, Inc. · Jun 24, 2003
- Modular Reach HipK994038 · Biomet, Inc. · Dec 22, 1999
- Biomet Modular Femoral Revision SystemK090757 · Biomet, Inc. · Sep 25, 2009
- Modular-Plus Revision StemK994126 · Plus Orthopedics · May 4, 2000
- Modular Plus Hip StemK032709 · Plus Orthopedics · Sep 23, 2003
Cited as a Predicate
3 later 510(k) summaries name this device as a predicate- EXPRT Revision Hip SystemK161610 · Encore Medical L.P. · Nov 9, 2016
- Revision Femoral StemK151739 · Lima Corporate S.P.A. · Nov 23, 2015
- Modulus Stems, Modulus Necks, Modular Femoral Heads, Cemented Cups, Lock Bipolar HeadsK112158 · Lima Corporate S.P.A. · May 10, 2012
Recalls
Recalls FDA ties to K092331, newest first; each opens FDA's recall record| Started | Reason | Status |
|---|---|---|
| Jan 10, 2017 | Z-1191-2017Lima Proximal Bodies were inadvertently re-sterilized. The safety screw which affixes the distal and proximal bodies of the stem includes a thread-locking plug made from UHMWPE, which is not approved for repeated gamma… | Terminated |
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Aug 4, 2009
- Review panel
- Orthopedic
- Regulation
- 21 CFR 888.3330
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files