Instrument, Coagulation, Automated. 510(k) K091980, from Instrumentation Laboratory CO (Bedford, MA). Product code GKP, class 2.

As of Nov 17, 2026, K091980 is an FDA 510(k) held by Instrumentation Laboratory CO, substantially equivalent on Sep 22, 2009 after 83 days, with 24 recalls tied to it in FDA's recall database.

  • Substantially equivalent: Sep 22, 2009
  • Review time: 83 days
  • Recalls: 24
  • Cited as a predicate by: 2 devices

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

The ACL TOP is a bench top, fully automated, random access analyzer designed specifically for in vitro diagnostic clinical use in the hemostasis laboratory for coagulation and/or fibrinolysis testing in the assessment of thrombosis and/or hemostasis. The system provides results for both direct hemostasis measurements and calculated parameters.

Predicate Devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a Predicate

2 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K091980, newest first; each opens FDA's recall record
Jun 3, 2014Z-2064-2014Potential for sample misidentification.Terminated
Jun 3, 2014Z-2062-2014Potential for sample misidentification.Terminated
Jun 3, 2014Z-2060-2014Potential for sample misidentification.Terminated
Jun 3, 2014Z-2065-2014Potential for sample misidentification.Terminated
Jun 3, 2014Z-2061-2014Potential for sample misidentification.Terminated
Jun 3, 2014Z-2063-2014Potential for sample misidentification.Terminated
Jun 3, 2014Z-2066-2014Potential for sample misidentification.Terminated
Jul 19, 2013Z-2245-2013Potential carryover issue that can can cause shortened APTT clotting times on the instrument.Terminated
Jul 19, 2013Z-2244-2013Potential carryover issue that can can cause shortened APTT clotting times on the instrument.Terminated
Jul 19, 2013Z-2246-2013Potential carryover issue that can can cause shortened APTT clotting times on the instrument.Terminated
Jul 19, 2013Z-2250-2013Potential carryover issue that can can cause shortened APTT clotting times on the instrument.Terminated
Jul 19, 2013Z-2247-2013Potential carryover issue that can can cause shortened APTT clotting times on the instrument.Terminated
Jul 19, 2013Z-2248-2013Potential carryover issue that can can cause shortened APTT clotting times on the instrument.Terminated
Jul 19, 2013Z-2249-2013Potential carryover issue that can can cause shortened APTT clotting times on the instrument.Terminated
Aug 3, 2012Z-2388-2012Low Abnormal Fibrinogen Sample Reports as Normal and High Fibrinogen Sample Reports as LowTerminated
Aug 3, 2012Z-2387-2012Low Abnormal Fibrinogen Sample Reports as Normal and High Fibrinogen Sample Reports as LowTerminated
Aug 3, 2012Z-2390-2012Low Abnormal Fibrinogen Sample Reports as Normal and High Fibrinogen Sample Reports as LowTerminated
Aug 3, 2012Z-2392-2012Low Abnormal Fibrinogen Sample Reports as Normal and High Fibrinogen Sample Reports as LowTerminated
Aug 3, 2012Z-2389-2012Low Abnormal Fibrinogen Sample Reports as Normal and High Fibrinogen Sample Reports as LowTerminated
Aug 3, 2012Z-2391-2012Low Abnormal Fibrinogen Sample Reports as Normal and High Fibrinogen Sample Reports as LowTerminated
Aug 13, 2010Z-0135-2011Software: New data check, "Multiple Threshold Check", was added to identify and fail abnormal (bimodal) clot curvesTerminated
Aug 13, 2010Z-0136-2011Software: New data check, "Multiple Threshold Check", was added to identify and fail abnormal (bimodal) clot curvesTerminated
Aug 13, 2010Z-0137-2011Software: New data check, "Multiple Threshold Check", was added to identify and fail abnormal (bimodal) clot curvesTerminated
Aug 13, 2010Z-0138-2011Software: New data check, "Multiple Threshold Check", was added to identify and fail abnormal (bimodal) clot curvesTerminated

Other GKP Clearances

The newest 510(k)s for the same product code; see every GKP device

Record

Type
Special
Received
Jul 1, 2009
Review panel
Hematology
Regulation
21 CFR 864.5400

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files