Philips NM3 Monitor, Model 7900
Spirometer, Monitoring (W/Wo Alarm). 510(k) K091459, from Respironics Novametrix, LLC (Wallingford, CT). Product code BZK, class 2.
As of Nov 18, 2026, K091459 is an FDA 510(k) held by Respironics Novametrix, LLC, substantially equivalent on Jul 28, 2009 after 71 days, with 4 recalls tied to it in FDA's recall database.
- Review time: 71 days
- Recalls: 4
- Cited as a predicate by: 1 device
Indications for use
As the 510(k) summary states them, read from FDA's summary PDFIntended Use: The intended use of the Philips NM3 monitor, Model 7900 is to provide: - cardiac output monitoring via the method of partial rebreathing in adult patients receiving mechanical ventilation during general anesthesia and in the intensive care unit (ICU).
Predicate devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- Nico With Mars, Model 7300K030886 · Respironics Novametrix, Inc. · Oct 7, 2003
Cited as a predicate
1 later 510(k) summary names this device as a predicate- Philips NM3 Respiratory Monitor With VentassistK103578 · Respironics Novametrix, LLC · Jun 2, 2011
Recalls
Recalls FDA ties to K091459, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record| Jul 29, 2011 | Z-3288-2011, Z-3287-2011Pressure lines maybe occluded and display incorrect tidal volumes. | Terminated |
| Jan 4, 2011 | Z-2168-2011, Z-2167-2011Software: two displayed respiratory parameters are incorrect due to errors in the system software: MValv (alveolar minute ventilation) and Vt/kg (tidal volume/patient weight). In both instances, the displayed parameters… | Terminated |
Other BZK clearances
The newest 510(k)s for the same product code; see every BZK device- ElintMAX Respiratory Monitor (ElintMAX Monitor A1)K260226 · Amptron Medical, Inc. · Sep 16, 2026
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Special
- Received
- May 18, 2009
- Review panel
- Anesthesiology
- Regulation
- 21 CFR 868.1850
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files