Aplio XG Diagnostic Ultrasound System Model SSA-790A Version 4.0
Transducer, Ultrasonic, Diagnostic. 510(k) K091295, from Toshiba America Medical Systems, In.C (Tustin, CA). Product code ITX, class 2.
As of Nov 21, 2026, K091295 is an FDA 510(k) held by Toshiba America Medical Systems, In.C, substantially equivalent on May 15, 2009 after 11 days, with 2 recalls tied to it in FDA's recall database.
- Review time: 11 days
- Recalls: 2
- Cited as a predicate by: 4 devices
- Aplio XG Diagnostic Ultrasound System, Model Ssa - 790A Version 3.0K082119 · Toshiba America Medical Systems, In.C · Aug 12, 2008
- Model Viamo SSA-640A, Version 2.0 Ultrasound SystemK100067 · Toshiba America Medical Systems, In.C · Jan 28, 2010
- Viamo SSA-640A Version 1.2K093171 · Toshiba America Medical Systems, In.C · Dec 4, 2009
- Aplio MX SSA-780A, Version 1.0K092948 · Toshiba America Medical Systems, In.C · Oct 9, 2009
- Aplio XG Diagnostic Ultrasound System, Model Ssa - 790A Version 4.0R001K092179 · Toshiba America Medical Systems, In.C · Jul 29, 2009
| Aug 5, 2013 | Z-0122-2014The firm initiated this recall of certain Toshiba Diagnostic Ultrasound Systems because it has been found that under certain conditions, the temperature of the transducer head may increase in 2D+PW mode or 2D+CDI… | Terminated |
| Jun 2, 2010 | Z-2981-2011Toshiba America Medical Systems Inc (TAMS) is initiating a field correction on the the Aplio XG model SSA-790A because it did not meet the requirement of clause 15b, IEC60601-1 which requires that the residual voltage… | Terminated |
- Align® Cover and Pad Guidance ToolK260114 · CIVCO Medical Instruments Co., Inc. · Sep 24, 2026
- BA-004CCK261170 · Fujifilm Corporation · Aug 12, 2026
- EM-TECH Passive Ultrasound Probe StrapK254143 · EM-TECH Solutions, Inc. · Aug 10, 2026
- SmartGuide Disposable Guides and GridsK260161 · UC-CARE , Ltd. · Jul 30, 2026
- ND Needle Guidance SystemK262474 · Accurate Access, LLC · Jul 27, 2026
- Accuro® 3S Needle Guide KitK261014 · Rivanna Medical, Inc. · Jul 24, 2026
- Ultrasonic Probe UM-3R (UM-3R); Ultrasonic Probe UM-G20-29R (UM-G20-29R)K250883 · Olympus Medical Systems Corporation · Sep 18, 2025
- Ultrasonic Probe UM-S20-17S (UM-S20-17S); Ultrasonic Probe UM-S20-20R (UM-S20-20R)K250762 · Olympus Medical Systems Corporation · Jul 11, 2025
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional, third-party review
- Received
- May 4, 2009
- Review panel
- Radiology
- Regulation
- 21 CFR 892.1570
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files