Sursigns VS2 Vital Signs Monitor, Suresigns VS3 Vital Signs Monitor, Suresigns VM Series Patient Monitors
Electrocardiograph. 510(k) K090483, from Philips Medical Systems (Andover, MA). Product code DPS, class 2.
As of Nov 19, 2026, K090483 is an FDA 510(k) held by Philips Medical Systems, substantially equivalent on Mar 20, 2009 after 23 days, with 5 recalls tied to it in FDA's recall database.
- Review time: 23 days
- Recalls: 5
- Cited as a predicate by: 6 devices
Indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients. Intended Use: For monitoring, recording, and alarming of multiple physiological parameters of adults, pediatrics and neonates in healthcare environments. Additionally, the monitor is intended for use in transport situations within a healthcare facility.
- M3046B Compact Portable Patient MonitorsK052707 · Philips Medical Systems · Oct 20, 2005
- Piconibp Oem ModuleK051366 · Philips Medizin Systeme Boeblingen GmbH · Sep 14, 2005
- Suresigns VM4, Suresigns VM6, Suresigns VM8K123900 · Philips Medical Systems · Apr 18, 2013
- Suresigns VS3 Vital Signs Monitor, Suresigns VS4 Vital Signs MonitorK120132 · Philips Medical Systems · Feb 14, 2012
- Suresigns VS2 Vital Signs MonitorK111114 · Philips Medical Systems · May 20, 2011
- Suresigns Vsi Vital Signs MonitorK110803 · Philips Medical Systems · May 20, 2011
- Suresigns VM Series Patient Monitors, Suresigns VM4, Suresigns VM6, and Suresigns VM8K103652 · Philips Medical Systesms · Jan 13, 2011
- Suresigns VM4, Suresigns VM6, Suresigns VM8 Model: 863063, 863064, 863065, 863066, 863068K101067 · Philips Medical Systems · May 7, 2010
| May 19, 2020 | Z-2337-2020Philips has identified a Philips SureSigns VM1 monitor in the possession of a clinical user that was not intended for commercial distribution and for which Philips does not have records regarding its manufacture.… | Terminated |
| Jul 13, 2018 | Z-2896-2018Philips has received several reports in which monitors with lithium ion batteries that had exceeded their battery life expectancy have overheated or ignited. These batteries should be replaced every 3 years or upon… | Open |
| Jun 10, 2011 | Z-2584-2011Speakers on the SureSigns Patient Monitors and Vital Signs, failure may be identified by the display of a Speaker Malfunc message in the alarm message area or an AUDIO FAILED message in the main screen of the device or… | Terminated |
| Oct 30, 2009 | Z-0374-2010, Z-0372-2010Use of these devices at their maximum volume setting may result in the premature failure of the internal speaker | Terminated |
- Kinetic Software LibraryK260640 · Monebo Technologies, Inc. · Oct 23, 2026
- MyoVista wavECG (DG-200-B15-003)K260005 · HeartSciences, Inc. · Oct 5, 2026
- QCA 10 Electrocardiograph (QCARD-10); QCA 10W Electrocardiograph (QCARD-10W)K260250 · QT Medical, Inc. · Sep 29, 2026
- SE-1202 Series Electrocardiograph (Model: SE-1202, SE-1202E)K252463 · Edan Instruments, Inc. · Jul 2, 2026
- MAC 7 Resting ECG Analysis SystemK251670 · Ge Medical Systems Information Technologies, Inc. · Jan 26, 2026
- AccurECG Analysis System (v2.0)K252361 · Accurkardia, Inc. · Dec 22, 2025
- Withings BeamO (SCT02)K252474 · Withings · Nov 10, 2025
- Cardiologs Holter PlatformK250569 · Philips France Commercial · Aug 6, 2025
- Document
- 510(k) summary (PDF) ↗
- Type
- Special
- Received
- Feb 25, 2009
- Review panel
- Cardiovascular
- Regulation
- 21 CFR 870.2340
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files