Arca
Skip to content
Tools
Tools

Clearvue Prone Breast Treatment Table

Accelerator, Linear, Medical. 510(k) K090349, from Orbital Therapy, LLC (Winchester, MA). Product code IYE, class 2.

As of Nov 21, 2026, K090349 is an FDA 510(k) held by Orbital Therapy, LLC, substantially equivalent on Mar 9, 2009 after 26 days, with 1 recall tied to it in FDA's recall database.

  • Review time: 26 days
  • Recalls: 1
  • Cited as a predicate by: 1 device

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

ClearVue, a breast radiation therapy patient support, is an accessory for medical linear accelerator radiation therapy systems. ClearVue is intended to be used to position and reposition patients for breast radiation therapy. ClearVue may be used with CT and MR scanners to acquire images for radiation therapy planning.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

1 later 510(k) summary names this device as a predicate
  • PronexK102765 · Direx Systems Corp. · Dec 7, 2010

Recalls

Recalls FDA ties to K090349, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
StartedRecallStatus
Jan 7, 2016Z-1264-2016Some of the Access ClearVue Prone Breast devices may contain scales that are misaligned from the zero reference with respect to other Access ClearVue Prone Breast devices. This can result in differences in scale…Terminated

Other IYE clearances

The newest 510(k)s for the same product code; their median review was 173 days; see every IYE device

Record

Type
Traditional
Received
Feb 11, 2009
Review panel
Radiology
Regulation
21 CFR 892.5050

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files