Hemosil Liquid Heparin, Heparin Calibrators And Lmw And Uf Heparin Controls
Assay, Heparin. 510(k) K090209, from Instrumentation Laboratory CO (Bedford, MA). Product code KFF, class 2.
As of Nov 17, 2026, K090209 is an FDA 510(k) held by Instrumentation Laboratory CO, substantially equivalent on Jun 2, 2009 after 125 days, with 5 recalls tied to it in FDA's recall database.
- Substantially equivalent: Jun 2, 2009
- Review time: 125 days
- Recalls: 5
- Cited as a predicate by: 2 devices
Indications for Use
As the 510(k) summary states them, read from FDA's summary PDFHemosIL Liquid Heparin: Automated chromogenic assay for the quantitative determination of unfractionated heparin (UFH) and low molecular weight heparin (LMWH) activity in human citrated plasma on IL Coagulation Systems (ACL TOP® Family, ACL™ ELITE/ELITE PRO®/8/9/10000 and ACL Futura/ACL Advance Systems).
Predicate Devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- Il Test HeparinK980242 · Instrumentation Laboratory CO · Mar 19, 1998
- Calibration Plasma Lmw HeparinK030964 · Instrumentation Laboratory CO · Jun 6, 2003
- Control Plasma Lmw HeparinK030965 · Instrumentation Laboratory CO · Jun 6, 2003
Cited as a Predicate
2 later 510(k) summaries name this device as a predicate- HemosIL Liquid Anti-XaK213464 · Instrumentation Laboratory CO · Oct 4, 2022
- VALIDATE Heparin Calibration Verification/Linearity Test KitK163498 · Maine Standards Company, LLC · Mar 16, 2017
Recalls
Recalls FDA ties to K090209, newest first; each opens FDA's recall record| Feb 3, 2023 | Z-1306-2023Removes the test definition for HemosIL Liquid Anti-Xa (PN 0020302600 and 0020302601) from the ACL Elite/Elite Pro instruments | Open, Classified |
| Feb 3, 2023 | Z-1305-2023Removes the test definition for HemosIL Liquid Anti-Xa (PN 0020302600 and 0020302601) from the ACL Elite/Elite Pro instruments | Open, Classified |
| Feb 3, 2023 | Z-1307-2023Removes the test definition for HemosIL Liquid Anti-Xa (PN 0020302600 and 0020302601) from the ACL Elite/Elite Pro instruments | Open, Classified |
| Feb 3, 2023 | Z-1304-2023Removes the test definition for HemosIL Liquid Anti-Xa (PN 0020302600 and 0020302601) from the ACL Elite/Elite Pro instruments | Open, Classified |
| Aug 25, 2021 | Z-0149-2022Labeled On-board instrument stability issue for current and future lots, reduced On-board Instrument Stability from 7 days to 5 days | Open, Classified |
Other KFF Clearances
The newest 510(k)s for the same product code; see every KFF device- STA® - Liquid Anti-XaK260214 · Diagnostica Stago, Inc. · Sep 24, 2026
- HemosIL Liquid Anti-XaK213464 · Instrumentation Laboratory CO · Oct 4, 2022
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Jan 28, 2009
- Review panel
- Hematology
- Regulation
- 21 CFR 864.7525
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files