Levitronix, Thoratec Centrimag Back-Up Consoles
Control, Pump Speed, Cardiopulmonary Bypass. 510(k) K090004, from Levitronix, LLC (Waltham, MA). Product code DWA, class 2.
As of Nov 20, 2026, K090004 is an FDA 510(k) held by Levitronix, LLC, substantially equivalent on Jan 15, 2009 after 13 days, with 1 recall tied to it in FDA's recall database.
- Review time: 13 days
- Recalls: 1
- Cited as a predicate by: 1 device
The Levitronix CentriMag Back-Up Console is indicated for use with the Levitronix CentriMag Extracorporeal Blood Pumping System. The Levitronix CentriMag Extracorporeal Blood Pumping System is indicated to pump blood through the extracorporeal bypass circuit for extracorporeal circulatory support for periods appropriate to cardiopulmonary bypass (up to six hours).
- Levitronix Centrimag Back-Up ConsoleK051209 · Levitronix, LLC · Sep 8, 2005
- Levitronix Centrimag Primary and Backup ConsolesK093832 · Levitronix, LLC · Jan 13, 2010
| Started | Recall | Status |
|---|---|---|
| Sep 5, 2018 | Z-0103-2019Reports of circulatory support system motor stopping. Motor stop can be attributed to a break of a wire or short circuit between wires within the electrical cable where the cable exits the motor cable bend protection. | Terminated |
- Inspira ART100K232788 · Inspira Technologies Oxy B.H.N. , Ltd. · May 24, 2024
- Anivia SG1000 Pump ConsoleK230698 · Apmtd, Inc. · Apr 18, 2023
- Anivia SG1000 Pump ConsoleK221491 · Apmtd, Inc. · Feb 3, 2023
- Document
- 510(k) summary (PDF) ↗
- Type
- Special
- Received
- Jan 2, 2009
- Review panel
- Cardiovascular
- Regulation
- 21 CFR 870.4380
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files