Interventional Fluoroscopic X-Ray System. 510(k) K083545, from Ziehm Imaging, Inc. (Riverside, CA). Product code OWB, class 2.

As of Nov 17, 2026, K083545 is an FDA 510(k) held by Ziehm Imaging, Inc., substantially equivalent on May 15, 2009 after 168 days, with 3 recalls tied to it in FDA's recall database.

  • Substantially equivalent: May 15, 2009
  • Review time: 168 days
  • Recalls: 3
  • Cited as a predicate by: 2 devices

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

The ZIEHM VISION RFD is intended for use in providing medical imaging, using pulsed and continuous fluoroscopic digital imaging, as well as digital subtraction and cine image capture during diagnostic interventional, and surgical procedures where intra-operative imaging and visualization of complex anatomical structures of both lower and higher contrast density are required, such procedures may…

Predicate Devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a Predicate

2 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K083545, newest first; each opens FDA's recall record
Apr 15, 2016Z-2641-2016Malfunction of the radiation-warning timer and the 3D-specific interval warning. No 5-minute timer dialogs or beeper will occur until the unit is rebooted.Terminated
Nov 21, 2013Z-0568-2014Engineering change in 2006 resulted in a minor non-conformity of their devices. Specifically the emergency switch was changed from a standard component to just an option on the devices.Terminated
Apr 29, 2011Z-1430-2012On 04/29/2011, Ziehm Imaging, Inc., Orlando, FL recalled the double foot switch, type KF 2 1S/1S-MED-AP, a component used with Ziehm Imaging GmbH Mobile C-arm due to the potential of unintentional continued activation…Terminated

Other OWB Clearances

The newest 510(k)s for the same product code; their median review was 231 days; see every OWB device

Record

Type
Traditional
Received
Nov 28, 2008
Review panel
Radiology
Regulation
21 CFR 892.1650

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Logos are each company's own site icon, matched by name through Wikidata.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.