Trilogy 100 Ventilator

Ventilator, Continuous, Facility Use. 510(k) K083526, from Respironics Inc., Sleep & Home Respiratory Group (Murrysville, PA). Product code CBK, class 2.

As of Nov 19, 2026, K083526 is an FDA 510(k) held by Respironics Inc., Sleep & Home Respiratory Group, substantially equivalent on Mar 13, 2009 after 105 days, with 8 recalls tied to it in FDA's recall database.

  • Review time: 105 days
  • Recalls: 8
  • Cited as a predicate by: 5 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

_____ Device Name: _ Trilogy 100 Ventilator The Respironics Trilogy100 system provides continuous or intermittent ventilatory support for the care of individuals who require mechanical ventilation. Trilogy100 is intended for pediatric through adult patients weighing at least 5 kg (11 lbs.).

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

5 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K083526, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Dec 7, 2022Z-0882-2023, Z-0884-2023The replacement silicone sound abatement foam installed into some Trilogy 100 and Trilogy 200 devices may separate from the plastic backing to which it is attached. If this were to happen, the foam could potentially…Open
Jun 14, 2021Z-1956-2021The polyester-based polyurethane (PE-PUR) foam may degrade into particles which may enter the devices air pathway and be ingested or inhaled by the user, and the PE-PUR foam may off-gas certain chemicals.Open
Jan 18, 2016Z-1065-2016Software IssueTerminated
Feb 11, 2014Z-1143-2014Internal testing revealed a potentially defective component on the power management board that could affect the function of certain Trilogy Ventilators resulting in loss of therapy to patients.Terminated
Apr 27, 2012Z-2024-2012The specified Trilogy units are being recalled because of the potential for a faulty component in the power supply.Terminated
Oct 11, 2011Z-0296-2012Trilogy 100, 200 and 202 units are being recalled due to low impeller press-force used during the manufacturing process. If the press force were low enough, theoretically the impeller could move up the shaft and come…Terminated
Jun 10, 2010Z-1896-2010It is possible that the battery could electrically short, resulting in a potential for fire or a thermal event.Terminated

Other CBK clearances

The newest 510(k)s for the same product code; their median review was 235 days; see every CBK device

Record

Type
Abbreviated
Received
Nov 28, 2008
Review panel
Anesthesiology
Regulation
21 CFR 868.5895

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files