Arca
Skip to content
Tools
Tools

Dialog With Adimea Option

Dialyzer, High Permeability With Or Without Sealed Dialysate System. 510(k) K083460, from B. Braun Avitum AG (Allentown, PA). Product code KDI, class 2.

As of Nov 19, 2026, K083460 is an FDA 510(k) held by B. Braun Avitum AG, substantially equivalent on Feb 18, 2010 after 454 days, with 8 recalls tied to it in FDA's recall database.

  • Review time: 454 days
  • Recalls: 8
  • Cited as a predicate by: 2 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

This dialysis machine can be used for implementing and monitoring hemodialysis treatments for patients with acute or chronic kidney failure. The system can be used in hospital, health center and outpatient dialysis center settings when prescribed by a physician.

Cited as a predicate

2 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K083460, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Jan 28, 2022Z-0872-2022, Z-0873-2022Firm learned of the potential for cracks to form in the conductivity sensors during use, which may lead to a leak in the dialysate circuit. This could result in excessive ultrafiltration or insufficient ultrafiltration…Terminated
Apr 1, 2016Z-1536-2016, Z-1531-2016, Z-1532-2016, Z-1535-2016, Z-1534-2016, Z-1533-2016Potential leakage of conductivity sensors of the Dialog Dialysis machines.Terminated

Other KDI clearances

The newest 510(k)s for the same product code; their median review was 189 days; see every KDI device

Record

Type
Traditional
Received
Nov 21, 2008
Review panel
Gastroenterology, Urology
Regulation
21 CFR 876.5860

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Esc

Type a name: a tool, a company, a drug or brand, a condition, a gene or a state. Identifiers like K231234, NCT05012345 or NDA 215866 open their record.