Skip to content
Tools
Tools

Habib Endohpb

Unit, Electrosurgical, Endoscopic (With Or Without Accessories). 510(k) K083292, from Emcision , Ltd. (London, Uk, GB). Product code KNS, class 2.

As of Nov 28, 2026, K083292 is an FDA 510(k) held by Emcision , Ltd., substantially equivalent on Jun 10, 2009 after 212 days, with 1 recall tied to it in FDA's recall database.

  • Review time: 212 days
  • Recalls: 1
  • Cited as a predicate by: 1 device

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

Indications For Use Statement 510(K) Number (if known) Not yet Allocated Device Name Habib EndoHPB The Habib EndoHPB is intended to be used to assist in the coagulation of tissue during Endoscopic surgical procedures in the Gastro-intestinal Tract. PLEASE DO NO WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

1 later 510(k) summary names this device as a predicate

Recalls

Recalls FDA ties to K083292, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
StartedRecallStatus
Feb 2, 2023Z-1397-2023Cease use of the Habib EndoHPB Bipolar RF Catheter with the ERBE VIO3 Generator immediately due to the potential for delivery of excessive energy leading to thermal injury or tissue damage. An inaccurate setting is…Open

Other KNS clearances

The newest 510(k)s for the same product code; their median review was 118 days; see every KNS device

Record

Type
Traditional, third-party review
Received
Nov 10, 2008
Review panel
Gastroenterology, Urology
Regulation
21 CFR 876.4300

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files