Automated External Defibrillators (Non-Wearable). 510(k) K082937, from Physio-Control, Inc. (Redmond, WA). Product code MKJ, class 3.

As of Nov 23, 2026, K082937 is an FDA 510(k) held by Physio-Control, Inc., substantially equivalent on Mar 11, 2009 after 161 days, with 3 recalls tied to it in FDA's recall database.

  • Review time: 161 days
  • Recalls: 3
  • Cited as a predicate by: 3 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

Manual Defibrillation Indications Manual defibrillation is indicated for the termination of certain potentially fatal arrhythmias, such as ventricular fibrillation and symptomatic ventricular tachycardia.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

3 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K082937, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Mar 11, 2010Z-0761-2017Customer complaint was received relating to a LifePak15 device that would power on then power down unexpectedly after a short period of use. When the LP15 powered down, power could be immediately restored by activation…Terminated
Mar 4, 2010Z-1410-2010Potential for the device to power off then on by itself, or to power off by itself and requiring the operator to turn it back on, or the device doesn't turn off.Terminated
Aug 12, 2009Z-2110-2009Some LIFEPAK 15 Monitor/Defibrillators have a Non-Invasive Blood Pressure (NIBP) module that was supplied by CAS Medical Systems, Branford, Connecticut who is recalling the modules. The NIBP modules may be susceptible…Terminated

Record

Type
Traditional
Received
Oct 1, 2008
Review panel
Cardiovascular
Regulation
21 CFR 870.5310

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files