INFX-8000H, Infinix-I

System, X-Ray, Fluoroscopic, Image-Intensified. 510(k) K082830, from Toshiba America Medical Systems, In.C (Tustin, CA). Product code JAA, class 2.

As of Nov 18, 2026, K082830 is an FDA 510(k) held by Toshiba America Medical Systems, In.C, substantially equivalent on Oct 10, 2008 after 14 days, with 2 recalls tied to it in FDA's recall database.

  • Substantially equivalent: Oct 10, 2008
  • Review time: 14 days
  • Recalls: 2
  • Cited as a predicate by: 1 device

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

This device is a digital radiography/fluoroscopy system used in a diagnostic angiography configuration. This system is indicated for use in diagnostic and angiographic procedures for blood vessels in the heart, brain, abdomen and lower extremities.

Predicate Devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a Predicate

1 later 510(k) summary names this device as a predicate

Recalls

Recalls FDA ties to K082830, newest first; each opens FDA's recall record
Dec 29, 2017Z-0845-2018During an examination a device error occurred and the ceiling drive operation was disabled. It was found that the fastening bolts of the driving portion of the ceiling drive loosened cause the error.Terminated
Oct 3, 2011Z-0331-2012The recall was initiated because Toshiba has confirmed that during patient transfer from a gurney to the catherization table, the tabletop may rotate unexpectedly when sufficient force is applied to the tabletop. The…Terminated

Other JAA Clearances

The newest 510(k)s for the same product code; their median review was 110 days; see every JAA device

Record

Type
Traditional, third-party review
Received
Sep 26, 2008
Review panel
Radiology
Regulation
21 CFR 892.1650

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files