Set, Administration, Intravascular. 510(k) K082806, from Icu Medical, Inc. (Salt Lake City, UT). Product code FPA, class 2.

As of Nov 20, 2026, K082806 is an FDA 510(k) held by Icu Medical, Inc., substantially equivalent on Apr 6, 2009 after 194 days, with 5 recalls tied to it in FDA's recall database.

  • Review time: 194 days
  • Recalls: 5
  • Cited as a predicate by: 2 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The Spiros is a single use, sterile, non-pyrogenic, swab-able, bidirectional valve device intended for use as an accessory to an Intravascular Administration Set. The Spiros provides access for the administration needle or catheter (which is inserted into the vein or artery).

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF
  • SpirosK070532 · Icu Medical, Inc. · Mar 20, 2007

Cited as a predicate

2 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K082806, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Aug 1, 2022Z-1723-2022Due to a manufacturing defect which results in a visible gap between the connector's top and bottom housing.Open
Mar 16, 2021Z-1569-2021, Z-1568-2021, Z-1567-2021Due to molding defect, there is a potential that certain lots of spinning and non-spinning Sprios male luers may exhibit small amounts of leakage in the inactivated position.Open
Aug 31, 2020Z-0702-2021Potential for spinning versions of male luers to exhibit leaks due to a molding defect. Fluid leakage may potentially cause delay of infusion, contamination of the fluid path, exposure to hazardous medications, blood…Open

Other FPA clearances

The newest 510(k)s for the same product code; their median review was 266 days; see every FPA device

Record

Type
Special
Received
Sep 24, 2008
Review panel
General Hospital
Regulation
21 CFR 880.5440

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files