Arca
Skip to content

Aequalis Humeral Nail System

Rod, Fixation, Intramedullary And Accessories. 510(k) K082754, from Tornier (Warsaw, IN). Product code HSB, class 2.

As of Nov 17, 2026, K082754 is an FDA 510(k) held by Tornier, substantially equivalent on Jan 27, 2009 after 130 days, with 2 recalls tied to it in FDA's recall database.

  • Substantially equivalent: Jan 27, 2009
  • Review time: 130 days
  • Recalls: 2
  • Cited as a predicate by: 2 devices

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

The Aequalis Humeral Nail System is intended to provide temporary stabilization of various types of proximal and/or diaphyseal fractures of the humerus. Types of fractures include, but are not limited to, non-unions, malunions, malalignments, pathological fractures, and impending pathological fractures.

Cited as a Predicate

2 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K082754, newest first; each opens FDA's recall record
Mar 26, 2018Z-1757-2018Drill bit tips without new BargerGard tip protector may puncture through sterile packaging.Terminated
Jun 25, 2015Z-2315-2015Recall for the Aequalis IM Nail instrumentation set (Tray Number 9020000) due to several reports that the mast of the targeting jig (Part Number 9020060) is separating from the jig boom. Separation will impact alignment…Terminated

Other HSB Clearances

The newest 510(k)s for the same product code; their median review was 192 days; see every HSB device

Record

Type
Traditional
Received
Sep 19, 2008
Review panel
Orthopedic
Regulation
21 CFR 888.3020

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Logos are each company's own site icon, matched by name through Wikidata.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.