Arca
Skip to content
Tools
Tools

Synthes Large External Fixation, Mr Conditional

Appliance, Fixation, Nail/Blade/Plate Combination, Multiple Component. 510(k) K082650, from Synthes (Usa) (West Chester, PA). Product code KTT, class 2.

As of Nov 19, 2026, K082650 is an FDA 510(k) held by Synthes (Usa), substantially equivalent on Nov 18, 2008 after 67 days, with 5 recalls tied to it in FDA's recall database.

  • Review time: 67 days
  • Recalls: 5
  • Cited as a predicate by: 8 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

Synthes Large External Fixation is intended for use to provide treatment for long bone and pelvic fractures that require external fixation. Specifically, the components can be used for: - Stabilization of soft tissues and fractures - Polytrauma/multiple orthopedic trauma - Vertically stable pelvic fractures, or as a treatment adjunct for vertically unstable pelvic fractures - Arthrodeses and…

Cited as a predicate

8 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K082650, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Jul 29, 2015Z-2451-2015Labeling changes were made related to MR conditions that align with standard definitions for MR Safe, MR Conditional, and MR Unsafe. This recall identifies additional Trauma External Fixation System part numbers that…Terminated
Apr 10, 2014Z-1833-2014Labeling changes have been made related to MR (Magnetic Resonance imaging) conditions as a result of changes in required protocols to designate a product MR Safe, MR Conditional, or MR Unsafe.Terminated
Jan 11, 2013Z-0941-2013The L1 hex nuts, used to tighten the Large Combination Clamp, were oversized and the 11 mm sockets and wrenches will not fit on the hex nuts.Terminated
Oct 12, 2012Z-1816-2013It was discovered internally that there was a discrepancy between the insert provided with the Tube to Tube Clamp and what was called for in the Bill of Materials. Following a detailed investigation, it was determined…Open
Sep 28, 2012Z-0277-2013The insert provided with the product for the specific lot number indicates that this product is MR safe. This product is MR Conditional and is etched as MR Conditional. The insert is incorrect.Terminated

Other KTT clearances

The newest 510(k)s for the same product code; their median review was 84 days; see every KTT device

Record

Type
Traditional
Received
Sep 12, 2008
Review panel
Orthopedic
Regulation
21 CFR 888.3030

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Esc

Type a name: a tool, a company, a drug or brand, a condition, a gene or a state. Identifiers like K231234, NCT05012345 or NDA 215866 open their record.

Logos are each company's own site icon, matched by name through Wikidata, or logos provided by Logo.dev.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.