Verathon Inc. Bladderscan Ultrasound System, Model Bvi 9600

System, Imaging, Pulsed Echo, Ultrasonic. 510(k) K082456, from Verathon Incorporated (Bothell, WA). Product code IYO, class 2.

As of Nov 19, 2026, K082456 is an FDA 510(k) held by Verathon Incorporated, substantially equivalent on Dec 1, 2008 after 97 days, with 3 recalls tied to it in FDA's recall database.

  • Review time: 97 days
  • Recalls: 3

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

Abdomen, B-Mode, per Indications for Use Ultrasound Form - The Verathon Inc. BladderScan® BVI 9600 with AortaScan® Mode is a user selectable, dual-function ultrasound device that projects ultrasound energy either into the lower abdomen to obtain an image of the bladder for measuring bladder volume, or into the mid-abdomen to obtain an image of the abdominal aorta for aortic diameter measurement.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Recalls

Recalls FDA ties to K082456, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Nov 30, 2015Z-1091-2016The firm is providing customers with an updated Operations and Maintenance Manual for each of the BladderScan BVI 9600 and AortaScan AMI 9700 devices. The updated Operations and Maintenance Manuals clarify that these…Terminated
Aug 23, 2010Z-0004-2011The AortaScan AMI 9700 devices may experience the loss of a factory installed software configuration file which results in a loss of the aorta measurement function.Terminated
Aug 23, 2010Z-0003-2011The BladderScan BVI 9600 with AortaScan devices may experience the loss of a factory installed software configuration file which results in a loss of the aorta measurement function.Terminated

Other IYO clearances

The newest 510(k)s for the same product code; their median review was 234 days; see every IYO device

Record

Type
Abbreviated
Received
Aug 26, 2008
Review panel
Radiology
Regulation
21 CFR 892.1560

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files