System, Image Processing, Radiological. 510(k) K082441, from Nordicimaginglab AS (Bergen, Hordaland, NO). Product code LLZ, class 2.

As of Nov 18, 2026, K082441 is an FDA 510(k) held by Nordicimaginglab AS, substantially equivalent on Nov 21, 2008 after 88 days, with 2 recalls tied to it in FDA's recall database.

  • Review time: 88 days
  • Recalls: 2
  • Cited as a predicate by: 1 device

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

nordiclCE is an image processing software package to be used by trained professionals including but not limited to physicians and medical technicians. The software runs on a standard "off-the-shelf" PC workstation and can be used to perform image viewing, processing and analysis of medical images. Data and images are acquired through DICOM compliant imaging devices and modalities.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

1 later 510(k) summary names this device as a predicate

Recalls

Recalls FDA ties to K082441, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Jul 3, 2015Z-2061-2018The bug causes BOLD activation maps to be visualized as overlays without taking the coregistration into account. Any output created from these activation maps as overlays will not be adjusted according to the…Terminated
Sep 12, 2014Z-2044-2018An error was discovered in the interpretation of certain DICOM header tags that may lead to incorrect orientation labeling, and thus and indirect left-right, up-down or anterior-posterior flipping of images.Terminated

Other LLZ clearances

The newest 510(k)s for the same product code; their median review was 145 days; see every LLZ device

Record

Type
Traditional
Received
Aug 25, 2008
Review panel
Radiology
Regulation
21 CFR 892.2050

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files