Nordicice Software
System, Image Processing, Radiological. 510(k) K082441, from Nordicimaginglab AS (Bergen, Hordaland, NO). Product code LLZ, class 2.
As of Nov 18, 2026, K082441 is an FDA 510(k) held by Nordicimaginglab AS, substantially equivalent on Nov 21, 2008 after 88 days, with 2 recalls tied to it in FDA's recall database.
- Review time: 88 days
- Recalls: 2
- Cited as a predicate by: 1 device
Indications for use
As the 510(k) summary states them, read from FDA's summary PDFnordiclCE is an image processing software package to be used by trained professionals including but not limited to physicians and medical technicians. The software runs on a standard "off-the-shelf" PC workstation and can be used to perform image viewing, processing and analysis of medical images. Data and images are acquired through DICOM compliant imaging devices and modalities.
Predicate devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- Nordic Image Control and Evaluation (Nice) Software, Version 2.1K063539 · Nordicicemedical AS · Jan 17, 2007
Cited as a predicate
1 later 510(k) summary names this device as a predicate- Modification To Nordicice SoftwareK090546 · Nordicimaginglab AS · Jun 8, 2009
Recalls
Recalls FDA ties to K082441, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record| Jul 3, 2015 | Z-2061-2018The bug causes BOLD activation maps to be visualized as overlays without taking the coregistration into account. Any output created from these activation maps as overlays will not be adjusted according to the… | Terminated |
| Sep 12, 2014 | Z-2044-2018An error was discovered in the interpretation of certain DICOM header tags that may lead to incorrect orientation labeling, and thus and indirect left-right, up-down or anterior-posterior flipping of images. | Terminated |
Other LLZ clearances
The newest 510(k)s for the same product code; their median review was 145 days; see every LLZ device- 3D Anatomical ModelK262132 · The Jacobs Institute · Nov 5, 2026
- LiLi DICOM Viewer (LiLi )K260454 · Elevate Imaging, Inc. · Nov 4, 2026
- Vitrea Advanced Visualization, Version 7.16K263648 · Canon Medical Informatics, Inc. · Oct 29, 2026
- MIM – KineticIDK261763 · Mim Software, Inc. · Oct 26, 2026
- PV-iCurve ProK261531 · Perception Vision Medical Technology Co., Ltd. · Oct 21, 2026
- Eden PACSK261788 · Higia., Inc. d/b/a Eden · Oct 21, 2026
- Syngo Carbon Space (VA4)K263048 · Siemens Healthineers AG · Sep 29, 2026
- Syneris PACS v.1.0.1.0K261874 · Syneris · Sep 23, 2026
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Aug 25, 2008
- Review panel
- Radiology
- Regulation
- 21 CFR 892.2050
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files