Syngo Imaging XS
System, Image Processing, Radiological. 510(k) K082430, from Siemens AG Medical Solutions (Erlangen, DE). Product code LLZ, class 2.
As of Nov 19, 2026, K082430 is an FDA 510(k) held by Siemens AG Medical Solutions, substantially equivalent on Nov 14, 2008 after 84 days, with 12 recalls tied to it in FDA's recall database.
- Review time: 84 days
- Recalls: 12
- Cited as a predicate by: 1 device
syngo Imaging XS is a Picture Archiving and Communication System (PACS) intended to display, process, read, report, communicate, distribute, store and archive digital medical images. It supports the physician in diagnosis and treatment planning. syngo Imaging XS also supports storage and archiving of Structured DICOM Reports.
- Syngo Imaging (Version V30)K071114 · Siemens AG Medical Solutions · May 22, 2007
- Visage Pacs/Cs, Release Version 5.0K072205 · Mercury Computer Systems, Inc. · Aug 29, 2007
- Syngo.PlazaK093612 · Siemens AG Medical Solutions · Mar 4, 2010
| Nov 17, 2015 | Z-0550-2016For the Syngo Workflow SLR System with Software Ver: VB10C: Printouts may be printed in incorrect anatomical size when using syngo Imaging XS filming application in conjunction with a printer not released for anatomical… | Terminated |
| Nov 11, 2014 | Z-0576-2015Printouts may be printed in incorrect anatomical size when using syngo.plaza or syngo Imaging XS filming application in conjunction with a printer not released for anatomical print usage. This may happen due to an… | Terminated |
| Jun 30, 2014 | Z-2111-2014Siemens has become aware of the potential malfunction when using syngo Imaging XS, version VA70A or higher. In some cases a miscalculation may occur for the grey scale values in the functions Region of Interest, Pixel… | Terminated |
| Jul 18, 2012 | Z-2264-2012There is a potential malfunction when using syngo Imaging XS, version VA70A or higher. In some situations it can happen that a miscalculation is done for the grey scale values in the functions "Regions of Interest",… | Terminated |
| Jan 26, 2012 | Z-1064-2012In September 2011 firm issued two Customer Safety Advisory Notices for syngo Imaging XS. Firm became aware of an unintended behavior when using syngo Imaging XS, version VA70A and VA70B with Client/Server Installations.… | Terminated |
| Sep 20, 2011 | Z-0514-2012Unintended behavior when using syngo Imaging XS, all versions. The "Rename" functionality can cause an unintended patient merge if the renamed patient is sent to another system. | Terminated |
| Sep 19, 2011 | Z-0515-2012Unintended behavior when using syngo Imaging XS, versions VA70A and VA70B Client/Serve installations. Loading additional images/series from Patient Browser using the thumbnails can cause an unintended deletion of… | Terminated |
| Apr 6, 2010 | Z-1892-2010Siemens has become aware of a potential malfunction and hazard to patients when using Syngo Imaging version lines VB20, VB30, VB35A in combination with a RIS that is violating DICOM Standard by creating non-unique study… | Terminated |
| Feb 16, 2010 | Z-1423-2010images can be overwritten | Terminated |
| Sep 30, 2009 | Z-0174-2010Incomplete amount of image data; specifically, The image count in the series table of the online and archive database will display the full amount of images received at the Syngo Imaging XS as archived. However, the zip… | Terminated |
| Sep 29, 2009 | Z-0270-2010, Z-0271-2010Distorted image. The "Distorted" image cannot be used for diagnostic findings, and, depending on the customer scenario, a repeated exam could be necessary if the image cannot be reported on any other workplace or… | Terminated |
- 3D Anatomical ModelK262132 · The Jacobs Institute · Nov 5, 2026
- LiLi DICOM Viewer (LiLi )K260454 · Elevate Imaging, Inc. · Nov 4, 2026
- Vitrea Advanced Visualization, Version 7.16K263648 · Canon Medical Informatics, Inc. · Oct 29, 2026
- MIM – KineticIDK261763 · Mim Software, Inc. · Oct 26, 2026
- PV-iCurve ProK261531 · Perception Vision Medical Technology Co., Ltd. · Oct 21, 2026
- Eden PACSK261788 · Higia., Inc. d/b/a Eden · Oct 21, 2026
- Syngo Carbon Space (VA4)K263048 · Siemens Healthineers AG · Sep 29, 2026
- Syneris PACS v.1.0.1.0K261874 · Syneris · Sep 23, 2026
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional, third-party review
- Received
- Aug 22, 2008
- Review panel
- Radiology
- Regulation
- 21 CFR 892.2050
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files