Arca
Skip to content
Tools
Tools

Iba Proton Therapy System- Proteus 235

System, Radiation Therapy, Charged-Particle, Medical. 510(k) K082416, from Ion Beam Applications S.A. (Philadelphia, PA). Product code LHN, class 2.

As of Nov 20, 2026, K082416 is an FDA 510(k) held by Ion Beam Applications S.A., substantially equivalent on Dec 12, 2008 after 113 days, with 3 recalls tied to it in FDA's recall database.

  • Review time: 113 days
  • Recalls: 3
  • Cited as a predicate by: 4 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The PTS is a medical device designed to product and deliver a proton beam for the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

4 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K082416, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Oct 19, 2017Z-0931-2018Range for the same energy could be different depending if the layer is the first layer of the irradiation plan or not.Terminated
Mar 27, 2014Z-0020-2015While the Proteus 235 is in the pencil beam scanning (PBS) treatment mode, on rare occasions, Low Level Radio Frequency (LLRF) pulsed dee voltage (i.e., voltage applied between two electrodes in the cyclotron to…Terminated
Dec 4, 2013Z-1492-2014Internal testing of the Proteus 235 Proton Therapy System revealed incorrect dose rate calculation in case of beam line option other than Zero (0).Terminated

Other LHN clearances

The newest 510(k)s for the same product code; their median review was 190 days; see every LHN device

Record

Type
Traditional
Received
Aug 21, 2008
Review panel
Radiology
Regulation
21 CFR 892.5050

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files