Arca
Skip to content
Tools
Tools

Sta Satellite Automated Multi-Parameter Analyzer

System, Multipurpose For In Vitro Coagulation Studies. 510(k) K082248, from Diagnostica Stago, Inc. (Parisppany, NJ). Product code JPA, class 2.

As of Nov 20, 2026, K082248 is an FDA 510(k) held by Diagnostica Stago, Inc., substantially equivalent on Dec 12, 2008 after 126 days, with 2 recalls tied to it in FDA's recall database.

  • Review time: 126 days
  • Recalls: 2
  • Cited as a predicate by: 1 device

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The Satellite™ is a fully automatic clinical instrument designed for the performance of tests on human plasmas, the results of which aid in the diagnosis of coagulation abnormalities or in monitoring anticoagulant therapy.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

1 later 510(k) summary names this device as a predicate

Recalls

Recalls FDA ties to K082248, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Jul 21, 2025Z-2473-2025Potential for contamination of STA-LIQUID ANTI-Xa by Stago Fibrinogen reagents in UFH and LMWH applications.Open
Jun 15, 2012Z-0252-2013Following a single customer complaint that incorrect INR values were occasionally printed for PT tests, even though the primary units (seconds) were correct, the firm investigated and confirmed the potential for…Terminated

Other JPA clearances

The newest 510(k)s for the same product code; their median review was 158 days; see every JPA device

Record

Type
Traditional
Received
Aug 8, 2008
Review panel
Hematology
Regulation
21 CFR 864.5425

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files