Amicas Pacs, Version 6.0

System, Image Processing, Radiological. 510(k) K082144, from Amicas, Inc. (Boston, MA). Product code LLZ, class 2.

As of Nov 19, 2026, K082144 is an FDA 510(k) held by Amicas, Inc., substantially equivalent on Sep 5, 2008 after 37 days, with 13 recalls tied to it in FDA's recall database.

  • Review time: 37 days
  • Recalls: 13
  • Cited as a predicate by: 2 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

AMICAS PACS 6.0 is designed and marketed for soft copy reading, communication and storage of studies produced by digital modalities, to include Digital Mammography. AMICAS PACS 6.0 receives images acquired from DICOM-compliant medical imaging systems, data from FDA-cleared Computer-Aided Detection systems and other FDA-cleared Image processing systems.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

2 later 510(k) summaries name this device as a predicate
  • Merge PACSK173475 · Merge Healthcare Incorporated · Dec 8, 2017
  • ImagequbeK124048 · Intuitive Imaging Informatics, LLC · Jan 29, 2013

Recalls

Recalls FDA ties to K082144, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
May 24, 2017Z-2601-2017Studies that have QC, PDE, or HL7 ADT updates applied may fail to propagate via telmed protocol to the destination server and may fail to import on the destination server.Terminated
May 8, 2017Z-2591-2017When cut lines are shown on an image, the lines may go from being vertical and then shift to horizontal causing confusion to the physician. This may cause a misdiagnosis.Terminated
May 8, 2017Z-2590-2017Merge PACS did not show unviewed images when the last view was skipped. There is a potential risk to health of a physician misdiagnosis because not all images are available for viewing.Terminated
Apr 4, 2016Z-1176-2017Potential exists for an incorrect patient image being displayed which could result in the delay in diagnosis or treatment.Terminated
Apr 4, 2016Z-1431-2017Studies that have QC, PDE, or HL7 ADT updates applied may fail to propagate via telmed protocol to the destination server and may fail to import on the destination server.Terminated
Mar 28, 2016Z-0726-2017Cut lines on the image may present horizontally rather than vertically.Terminated
Mar 28, 2016Z-0599-2017The software did not show unviewed images when the last view was skipped. The physician may potentially misdiagnose when not all images are available for viewing.Terminated
Feb 5, 2016Z-2363-2016When measuring a lesion on an unmagnified mammography image then performing the same measurement on an image magnified by the Hologic imager (different image), the measurements are not the same.Terminated
Jan 30, 2016Z-2379-2016The patient name in the Halo title bar and the thumbnails do not match the name on displayed images.Terminated
Jan 30, 2016Z-2301-2016Studies coming over via telmed were missing patient's DOB, procedure, and referring physician.Terminated
Jan 30, 2016Z-2303-2016A migrated study that has annotations will display the annotations, but when a prior study is loaded that should have annotations, the second study will not display annotations.Terminated
Jan 30, 2016Z-2532-2016, Z-2531-2016Studies that are viewed in iConnect Access that originate on Merge PACS / Ortho PACS may not display the correct current patient demographics if there has been a PDE (Patient / Study Demographic Edit) performed after…Terminated

Other LLZ clearances

The newest 510(k)s for the same product code; their median review was 145 days; see every LLZ device

Record

Type
Traditional
Received
Jul 30, 2008
Review panel
Radiology
Regulation
21 CFR 892.2050

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files