Nannolight Intense Pulsed Light System
Powered Laser Surgical Instrument. 510(k) K082033, from Global USA Distribution, LLC (Minneapolis, MN). Product code GEX, class 2.
As of Nov 17, 2026, K082033 is an FDA 510(k) held by Global USA Distribution, LLC, substantially equivalent on Jul 31, 2008 after 14 days, with 1 recall tied to it in FDA's recall database.
- Substantially equivalent: Jul 31, 2008
- Review time: 14 days
- Recalls: 1
- Cited as a predicate by: 1 device
Indications for Use
As the 510(k) summary states them, read from FDA's summary PDFThe Nannolight Intense Pulsed Light System is indicated for use in surgical, aesthetic and cosmetic applications requiring selective photothermolysis (photocoagulation or coagulation) and hemostasis of soft tissue in the medical specialties of general and plastic surgery and dermatology.
Predicate Devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- Profile Bbl SystemK032460 · Sciton, Inc · Nov 13, 2003
- Quantum, Models Sr, Hr, Dl; Vasculight, Models Sr, Hr, Vs, Vx, DlK020839 · Lumenis, Inc. · Oct 11, 2002
Cited as a Predicate
1 later 510(k) summary names this device as a predicate- Palomar LUX1540 HandpieceK090195 · Palomar Medical Technologies, Inc. · Nov 20, 2009
Recalls
Recalls FDA ties to K082033, newest first; each opens FDA's recall record| Started | Reason | Status |
|---|---|---|
| Apr 12, 2008 | Z-1444-2009Sybaritic Inc. had an issue with the 585nm filter used with the NannoLight MP50. The filter may cause unwanted burning and should not be used for vascular lesion treatments. | Terminated |
Other GEX Clearances
The newest 510(k)s for the same product code; their median review was 90 days; see every GEX device- B`orze Inc Laser Family (Sorren/Adonis)K263063 · Borze, Inc. · Oct 30, 2026
- Diode laser device (DT25-BI, DT212-BI, DT213-BI, DT214-BI, DT215-BI)K262349 · Risu Medical Technology Co., Ltd. · Oct 27, 2026
- Fractional CO2 Laser Therapy System (Z121)K262627 · Sanhe PERFECTLASER Technology Co., Ltd. · Oct 26, 2026
- PicoPeak Picosecond Nd:Yag Laser System (Bella-9)K262701 · Beijing Sea Heart International Science and Technologyco.Ltd · Oct 22, 2026
- Picosecond Laser treatment system (Z180)K262628 · Sanhe PERFECTLASER Technology Co., Ltd. · Oct 20, 2026
- Nd:YAG & Alexandrite Long Pulse Laser Therapy Instrument (LPDuo)K261572 · Keyue (Hangzhou) Medical Information Technology Co., Ltd. · Oct 13, 2026
- CureLight Medical Diode Laser Systems (S3-BF45);CureLight Medical Diode Laser Systems (S2-B30);CureLight Medical Diode…K262346 · Wuhan Phomed Technology Co., Ltd. · Oct 8, 2026
- Diode Laser Hair Removal Machine (QDTM-03)K261692 · Beijing Nubway S&T Co., Ltd. · Oct 7, 2026
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional, third-party review
- Received
- Jul 17, 2008
- Review panel
- General, Plastic Surgery
- Regulation
- 21 CFR 878.4810
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files