Arca
Skip to content
Tools
Tools

Accuflow and Accuflux Elastomeric Infusion Pump

Pump, Infusion, Elastomeric. 510(k) K081905, from Westmed, Inc. (Highlands Ranch, CO). Product code MEB, class 2.

As of Nov 19, 2026, K081905 is an FDA 510(k) held by Westmed, Inc., substantially equivalent on Jul 15, 2008 after 12 days, with 7 recalls tied to it in FDA's recall database.

  • Review time: 12 days
  • Recalls: 7
  • Cited as a predicate by: 1 device

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

AccuFlo & AccuFlux devices are intended for patients requiring intravenous, percutaneous, subcutaneous, intra-operative sites or epidural administration of medications. It is the responsibility of the user to ensure that the medication is prepared and administered accordance with the drug manufacturer's package insert.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

1 later 510(k) summary names this device as a predicate

Recalls

Recalls FDA ties to K081905, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Aug 6, 2014Z-2655-2014The primary packaging label (inner pouch label) indicates that the package contents are Catalog Item: CT-1000-250, while the secondary packaging label (outer carton label) indicates the contents are CT-1000-100. The…Terminated
Dec 13, 2013Z-0921-2014Internal testing of two lots of the AccuFlo CT-2000-100 Elastomeric Infusion Pump identified flow rates which are outside the specified infusion time and may result in overly rapid infusion of medications.Terminated
Oct 20, 2011Z-0268-2012, Z-0269-2012, Z-0267-2012Redi-Flo Elastomeric Infusion Pumps may have a higher flow rate than specified (more than +15%).Terminated
Jul 28, 2011Z-0081-2012The product may have a higher flow rate than specified resulting in over administration of drug solutions to the patients.Terminated
Jul 9, 2011Z-0161-2013Please be aware that this is not a new recall. The firm has taken action; but, due to administrative issues this recall is now being reclassified by the Agency as a Class I. The recall was initiated because First…Terminated

Other MEB clearances

The newest 510(k)s for the same product code; their median review was 245 days; see every MEB device

Record

Type
Traditional, third-party review
Received
Jul 3, 2008
Review panel
General Hospital
Regulation
21 CFR 880.5725

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Esc

Type a name: a tool, a company, a drug or brand, a condition, a gene or a state. Identifiers like K231234, NCT05012345 or NDA 215866 open their record.