Advia Chemistry Hemoglobin A1C, Advia Chemistry A1C Calibrators
Assay, Glycosylated Hemoglobin. 510(k) K081895, from Siemens Medical Solutions Diagnostics (Tarrytown, NY). Product code LCP, class 2.
As of Nov 22, 2026, K081895 is an FDA 510(k) held by Siemens Medical Solutions Diagnostics, substantially equivalent on Jan 8, 2009 after 189 days, with 2 recalls tied to it in FDA's recall database.
- Review time: 189 days
- Recalls: 2
- Cited as a predicate by: 1 device
The ADVIA ® Chemistry Hemoglobin A1c method is for in vitro diagnostic use in the quantitative determination of Hemoglobin A1c, a diabetes marker, in whole blood on the ADVIA Chemistry systems. Such measurements are used for monitoring the long-term glycemic control of persons with diabetes.
- G7 Automated Hplc AnalyzerK011434 · Tosoh Medics, Inc. · Sep 18, 2001
- The Vitros 5,1 FS Chemistry SystemK041764 · Ortho-Clinical Diagnostics · Sep 9, 2004
- Advia Chemistry Hemoglobin A1C_3 (A1C_3) Automated Pretreatment AssayK110934 · Siemens Healthcare Diagnostics · Aug 25, 2011
| Aug 6, 2014 | Z-2643-2014The ADVIA Chemistry Systems A1c_3 method may exhibit a percent bias of -9% to +11% for hemoglobin (Hb)A1c samples when using A1c_3 Calibrator lots 3HD044 and 3LD068. Depending on quality control limits, this issue may… | Terminated |
| Jul 30, 2013 | Z-2135-2013Siemens confirmed that the ADVIA Chemistry Systems HbA1c method exhibits a positive bias of up to 12% for patient and CAP survey samples due to over-recovery when using A1c_3 Calibrator lots 1MD014, 1BD063, and 2GD014.… | Terminated |
- Q-Pad Test SystemK231465 · Qurasense · Dec 6, 2023
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Jul 3, 2008
- Review panel
- Hematology
- Regulation
- 21 CFR 864.7470
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files