Ncb Polyaxial Locking Plate System, Proximal Humeral Plates, Zimmer Universal Locking System, 3.5MM Tivanium TI-6AL-4V
Plate, Fixation, Bone. 510(k) K081759, from Zimmer GmbH (Warsaw, IN). Product code HRS, class 2.
As of Nov 19, 2026, K081759 is an FDA 510(k) held by Zimmer GmbH, substantially equivalent on Oct 16, 2008 after 118 days, with 2 recalls tied to it in FDA's recall database.
- Review time: 118 days
- Recalls: 2
- Cited as a predicate by: 4 devices
The $NCB$ Polyaxial Locking Plate System is indicated for temporary internal fixation and stabilization of fractures and osteotomies of long bones.
- Ncb Plating SystemK042695 · Zimmer GmbH · Oct 29, 2004
- Universal Locking System, 3.5MM Plates and Screws-4936 Plates,4835 ScrewsK060710 · Zimmer, Inc. · Apr 26, 2006
- NCB® Polyaxial Locking Plate System; NCB® Periprosthetic Femur Plate System; NCB® Cable Button; NCB® Straight Narrow…K241754 · Zimmer Switzerland Manufacturing GmbH · Aug 15, 2024
- NCB Polyaxial Locking Plate SystemK192021 · Zimmer GmbH · Aug 22, 2019
- Proximal Humerus Plating SystemK141493 · Miami Device Solutions · Mar 20, 2015
- Motionloc Screw for Ncb Polyaxial Locking Plating SystemK101696 · Zimmer, Inc. · Sep 10, 2010
| Oct 10, 2019 | Z-0332-2020Non-highly polished implants may adhere to the LDPE bag (previous packaging configuration). | Terminated |
| Sep 4, 2012 | Z-0246-2013Zimmer Inc., is initiating a correction to the patient labels of products manufactured before March 20I0 by the Zimmer GmbH production site in Winterthur, Switzerland, after receiving four complaints of a discrepancy… | Terminated |
- HERO Sternal Closure SystemK262939 · Hero Orthopedics, LLC · Nov 2, 2026
- Arthrex Plates, Screws and Pins, and WashersK263602 · Arthrex, Inc. · Oct 29, 2026
- Trochlear Salvage Plating SystemK261135 · Skeletal Dynamics, Inc. · Oct 26, 2026
- ANTHEM™ Fracture SystemK262779 · Globus Medical, Inc. · Oct 23, 2026
- Tempo Plating SystemK263407 · Pace Surgical · Oct 23, 2026
- Park's Pectus SystemK263269 · Tdm Co., Ltd. · Oct 15, 2026
- TDM Large Bone Plate and Screw SystemK261611 · Tdm Co., Ltd. · Oct 2, 2026
- Alians ForearmK262649 · Newclip Technics · Sep 28, 2026
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Jun 20, 2008
- Review panel
- Orthopedic
- Regulation
- 21 CFR 888.3030
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files