Arca
Skip to content

Tiger Cannulated Screw System

Screw, Fixation, Bone. 510(k) K081510, from Trilliant Surgical, Ltd. (Round Rock, TX). Product code HWC, class 2.

As of Nov 17, 2026, K081510 is an FDA 510(k) held by Trilliant Surgical, Ltd., substantially equivalent on Aug 4, 2008 after 67 days, with 5 recalls tied to it in FDA's recall database.

  • Substantially equivalent: Aug 4, 2008
  • Review time: 67 days
  • Recalls: 5
  • Cited as a predicate by: 5 devices

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

The Tiger Cannulated Screw Fixation System implants (screws) are intended for fixation of fractures, non-unions, arthrodeses and osteotomies of the small bones in the hand and foot.

Cited as a Predicate

5 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K081510, newest first; each opens FDA's recall record
Jul 19, 2018Z-1605-2020Instructions for Use (IFU) and Surgical Cleaning and Sterilization Protocols updated with revised dry time parameters. This recall of IFU occurred in 2018.Terminated
Jul 20, 2015Z-2452-2015The cannulation of the 3.0/4.0 Cannulated Drill Bit inner diamter is too small to allow for the guide wire to pass completely through.Terminated
Jul 20, 2015Z-2421-2015The cutting efficiency of the Countersinks was out of specification.Terminated
Jul 20, 2015Z-2464-2015Sterilization tray lids for the Tiger and Tiger Headless Cannulated Screw Systems were reported to occasionally be difficult to remove from the sterilization tray.Terminated
Apr 14, 2009Z-1438-2010Firm received several complaints of field failures, e.g., "broke during surgery" and the firm initiated a design change. A communication to sales staff 04/14/2009 instructed them to recover and reconcile these products…Terminated

Other HWC Clearances

The newest 510(k)s for the same product code; their median review was 157 days; see every HWC device

Record

Type
Traditional
Received
May 29, 2008
Review panel
Orthopedic
Regulation
21 CFR 888.3040

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files