Cusa Selector Nxt Ultrasonic Tissue Ablation System
Instrument, Ultrasonic Surgical. 510(k) K081459, from Integra Radionics, Inc. (Burlington, MA). Product code LFL, class U.
As of Nov 20, 2026, K081459 is an FDA 510(k) held by Integra Radionics, Inc., substantially equivalent on Aug 13, 2008 after 82 days, with 4 recalls tied to it in FDA's recall database.
- Review time: 82 days
- Recalls: 4
- Cited as a predicate by: 4 devices
The CUSA Selector™ NXT Ultrasonic Surgical Aspirator System is indicated for use in surgical procedures where fragmentation, emulsification and aspiration of soft and hard (e.g. bone) tissue is desirable, including Neurosurgery, Gastrointestinal and affiliated organ surgery, Urological surgery, Plastic and Reconstructive surgery, General surgery, Orthopedic surgery, Gynecological surgery…
- Integra Selector Ultrasonic Surgical Aspirator System With Bone TipK071669 · Integra Radionics, Inc. · Aug 17, 2007
- Selector Quantum Ultrasonic Aspirator SystemK042277 · Integra Neurosciences, Ltd. · Sep 29, 2004
- Selector Integra Ultrasonic Surgical Aspirator, Model Om 1530000M3K021989 · Integra Neurosciences · Sep 13, 2002
- CUSA Clarity Ultrasonic Surgical Aspirator SystemK161882 · Integra LifeSciences Corporation · Oct 14, 2016
- Integra(Tm) Cusa Nxt(Tm) Bone TipsK111741 · Integra Burlington Ma, Inc. (Formerly Integra Radi · Aug 23, 2011
- Cusa Nxt Extended Length TipK102258 · Integra Radionics, Inc. · Jun 10, 2011
- Cusa Shear Tip, An Accessory of The Cusa Excel Ultrasonic Surgical Aspirator SystemK090864 · Integra Radionics, Inc. · May 8, 2009
| Sep 29, 2014 | Z-0491-2015Some specific lots of Selector Tubing may potentially leak irrigation fluid during use if not placed on the Selector handpiece in a specific manner. | Terminated |
| May 2, 2011 | Z-2585-2011The insulated covering of a wire cable that powers the vacuum pump in the CUSA NXT Service Module can become abraded over time resulting in the exposure of bare wire. If this bare wire were to come in contact with the… | Terminated |
| Jul 29, 2010 | Z-2637-2010, Z-2638-2010A ground fault malfunction at the rectifier retaining screw may cause the aspiration pump in the CUSA NXT Service module to run at maximum speed, and continue to run until the entire system is turned off. This may… | Terminated |
- TORUS Ultrasonic Aspirator System; and accessories (TORUS Single-Use Tubing Sets, TORUS Single-Use Torque Wrenches,…K261361 · Carl Zeiss Meditec, AG · Oct 15, 2026
- Sonopet iQ Series III Console (5700-050-000); Sonopet iQ+ Straight Handpiece (5500-260-000); Sonopet iQ+ 10mm iQ+ Blade…K262244 · Stryker Instruments · Oct 8, 2026
- neXus Ultrasonic Surgical Aspirator SystemK261448 · Misonix, LLC · Jul 30, 2026
- CUSA® Clarity Ultrasonic Surgical Aspirator SystemK251162 · Integra LifeSciences Corporation · Oct 8, 2025
- SanAgile™ Ultrasonic Surgery Advanced Portable Controller (SA10); SanAgile™ Ultrasonic Surgery Advanced Dissector…K242894 · Shanghai Saints Sages Surgical Co., Ltd. · Jun 18, 2025
- Sonopet iQ Ultrasonic Aspirator System (5500-050-000)K243930 · Stryker Instruments · Mar 24, 2025
- Tenex 2nd Generation SystemK241700 · Trice Medical, Inc. · Nov 18, 2024
- CUSA® Clarity Ultrasonic Surgical Aspirator SystemK240493 · Integra LifeSciences Corporation · Jul 11, 2024
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- May 23, 2008
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files