Chemoclave Cytotoxic Medication Preparation and Delivery System, Model, Ch-Xxxx

Set, Administration, Intravascular. 510(k) K081361, from Icu Medical, Inc. (Salt Lake City, UT). Product code FPA, class 2.

As of Nov 20, 2026, K081361 is an FDA 510(k) held by Icu Medical, Inc., substantially equivalent on Aug 28, 2008 after 105 days, with 7 recalls tied to it in FDA's recall database.

  • Review time: 105 days
  • Recalls: 7
  • Cited as a predicate by: 3 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The ChemoCLAVE Cytotoxic Medication Preparation and Delivery System consists of 6 previously cleared components (CLAVE®, Spikes, SPIROS™, GENIE™, Vial Access, and Admin Sets) that can be combined into various configurations intended for use in the preparation and patient administration of cytotoxic medications.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

3 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K081361, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
May 12, 2026Z-3160-2026Intravascular administration tubing set drip chambers may have black specks potentially due to the plastic molding process leaving discolored polyvinyl chloride (PVC) material within the drip chamber wall. The presence…Open
May 12, 2026Z-3158-2026Intravascular administration tubing set drip chambers may have black specks potentially due to the plastic molding process leaving discolored polyvinyl chloride (PVC) material within the drip chamber wall that does not…Open
Mar 16, 2021Z-1565-2021, Z-1566-2021Due to molding defect, there is a potential that certain lots of spinning and non-spinning Sprios male luers may exhibit small amounts of leakage in the inactivated position.Open
Aug 31, 2020Z-0701-2021Potential for spinning versions of male luers to exhibit leaks due to a molding defect. Fluid leakage may potentially cause delay of infusion, contamination of the fluid path, exposure to hazardous medications, blood…Open
Feb 21, 2019Z-1092-2019, Z-1093-2019There is a potential for certain lots of ChemoLock and ChemoClave Vial Spikes to have plastic burr particulate originating from the protective cap used in the assembly of the device.Terminated

Other FPA clearances

The newest 510(k)s for the same product code; their median review was 266 days; see every FPA device

Record

Type
Traditional
Received
May 15, 2008
Review panel
General Hospital
Regulation
21 CFR 880.5440

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files