Legion Hinge Knee System
Prosthesis, Knee, Femorotibial, Constrained, Cemented, Metal/Polymer. 510(k) K081111, from Smith & Nephew, Inc. (Memphis, TN). Product code KRO, class 2.
As of Nov 17, 2026, K081111 is an FDA 510(k) held by Smith & Nephew, Inc., substantially equivalent on Jul 23, 2008 after 96 days, with 7 recalls tied to it in FDA's recall database.
- Substantially equivalent: Jul 23, 2008
- Review time: 96 days
- Recalls: 7
- Cited as a predicate by: 2 devices
Indications for Use
As the 510(k) summary states them, read from FDA's summary PDF1. Rheumatoid arthritis. 2. Post-traumatic arthritis, osteoarthritis, or degenerative arthritis in older patients whose age, weight, and activity level are compatible with an adequate long-term result. 3. Failed osteotomies, unicomparternal replacement, or total knee replacement. 4.
Cited as a Predicate
2 later 510(k) summaries name this device as a predicate- LEGION Knee SystemK180334 · Smith & Nephew, Inc. · Mar 5, 2018
- Smith and Nephew Legion Hinge Knee SystemK151118 · Smith & Nephew · Jul 28, 2015
Recalls
Recalls FDA ties to K081111, newest first; each opens FDA's recall record| Feb 1, 2018 | Z-1105-2018The screws package with devices within lot 14AAP6352A were 5 MM instead of 10MM and are too short to be used with the Legion HK Distal Wedge Size 5. | Terminated |
| May 26, 2016 | Z-2093-2016The surface inside of the femoral component, where cement is applied for adhesion, is below specification. | Terminated |
| May 26, 2016 | Z-2091-2016The surface inside of the femoral component, where cement is applied for adhesion, is below specification. | Terminated |
| May 26, 2016 | Z-2090-2016The surface inside of the femoral component, where cement is applied for adhesion, is below specification. | Terminated |
| May 26, 2016 | Z-2094-2016The surface inside of the femoral component, where cement is applied for adhesion, is below specification. | Terminated |
| May 26, 2016 | Z-2095-2016The surface inside of the femoral component, where cement is applied for adhesion, is below specification. | Terminated |
| May 26, 2016 | Z-2092-2016The surface inside of the femoral component, where cement is applied for adhesion, is below specification. | Terminated |
Other KRO Clearances
The newest 510(k)s for the same product code; their median review was 87 days; see every KRO device- Triathlon® Knee System; Triathlon® Hinge Knee System; Scorpio® Total Knee SystemK261898 · Howmedica Osteonics Corp. dba Stryker Orthopaedics · Aug 6, 2026
- Triathlon® Tritanium® Asymmetric Patella; Triathlon® Tritanium® Symmetric Patella;…K252898 · Howmedica Osteonics Corp., Dba Stryker Orthopaedics · Dec 15, 2025
- Stem Extension Line (USTAR II System)K252303 · United Orthopedic Corporation · Aug 22, 2025
- Triathlon® Hinge Knee SystemK251665 · Howmedica Osteonics Corp., Dba Stryker Orthopaedics · Jul 25, 2025
- ATTUNE Revision Hinge KneeK242871 · Depuy Ireland UC · Nov 21, 2024
- EVOLUTION® Hinge Knee System;GUARDIAN limb Salvage System;DCW Modular Distal Femoral…K240043 · Microport Orthopedics, Inc. · Apr 1, 2024
- ATTUNE Total Knee System, ATTUNE Cementless Knee System, LPS Limb Preservation System,…K233980 · Depuy Ireland UC · Mar 21, 2024
- EVOLUTION® Hinge Knee SystemK230563 · Microport Orthopedics, Inc. · Sep 27, 2023
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Apr 18, 2008
- Review panel
- Orthopedic
- Regulation
- 21 CFR 888.3510
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files