Arca
Skip to content

Legion Hinge Knee System

Prosthesis, Knee, Femorotibial, Constrained, Cemented, Metal/Polymer. 510(k) K081111, from Smith & Nephew, Inc. (Memphis, TN). Product code KRO, class 2.

As of Nov 17, 2026, K081111 is an FDA 510(k) held by Smith & Nephew, Inc., substantially equivalent on Jul 23, 2008 after 96 days, with 7 recalls tied to it in FDA's recall database.

  • Substantially equivalent: Jul 23, 2008
  • Review time: 96 days
  • Recalls: 7
  • Cited as a predicate by: 2 devices

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

1. Rheumatoid arthritis. 2. Post-traumatic arthritis, osteoarthritis, or degenerative arthritis in older patients whose age, weight, and activity level are compatible with an adequate long-term result. 3. Failed osteotomies, unicomparternal replacement, or total knee replacement. 4.

Cited as a Predicate

2 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K081111, newest first; each opens FDA's recall record
Feb 1, 2018Z-1105-2018The screws package with devices within lot 14AAP6352A were 5 MM instead of 10MM and are too short to be used with the Legion HK Distal Wedge Size 5.Terminated
May 26, 2016Z-2093-2016The surface inside of the femoral component, where cement is applied for adhesion, is below specification.Terminated
May 26, 2016Z-2091-2016The surface inside of the femoral component, where cement is applied for adhesion, is below specification.Terminated
May 26, 2016Z-2090-2016The surface inside of the femoral component, where cement is applied for adhesion, is below specification.Terminated
May 26, 2016Z-2094-2016The surface inside of the femoral component, where cement is applied for adhesion, is below specification.Terminated
May 26, 2016Z-2095-2016The surface inside of the femoral component, where cement is applied for adhesion, is below specification.Terminated
May 26, 2016Z-2092-2016The surface inside of the femoral component, where cement is applied for adhesion, is below specification.Terminated

Other KRO Clearances

The newest 510(k)s for the same product code; their median review was 87 days; see every KRO device

Record

Type
Traditional
Received
Apr 18, 2008
Review panel
Orthopedic
Regulation
21 CFR 888.3510

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Logos are each company's own site icon, matched by name through Wikidata.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.